8 indexes, 1 explanation of the messages, 1 error messages – YSI IQ S ENSOR N ET MIQ/CHV PLUS User Manual

Page 37: 2 informative messages, Indexes -1, Explanation of the messages -1 8.1.1, Error messages -1, Informative messages -1, 8indexes

Advertising
background image

MIQ/CHV PLUS

Indexes

8 - 1

ba76036e01

01/2012

8

Indexes

8.1

Explanation of the messages

In this chapter you will find a list with all the message codes and
corresponding message texts that may occur in the log book of the
IQ S

ENSOR

N

ET

system for the MIQ/CHV PLUS output module.

Note
Information about

Contents and structure of the log book and

Structure of the message code

can be found in the L

OG

BOOK

chapter of the IQ S

ENSOR

N

ET

system

operating manual.

Note
All message codes of the MIQ/CHV PLUSoutput module end with the
number, "432".

8.1.1

Error messages

8.1.2

Informative messages

The MIQ/CHV PLUS output module does not send any informative
messages.

Message code

Message text

EI2432

Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Defective components,
replace components
* Defective components,
replace components

Advertising