Cleaning, Pta only) – YSI SensoLyt serries User Manual

Page 3

Advertising
background image

SensoLyt

®

combination electrodes

Cleaning

3

ba76005e01

01/2012

Cleaning

Conversion to standard hydrogen electrode

(SensoLyt

®

PtA only)

U

H

= U

Meas

+ U

Ref

U

Ref

depends on the reference system and temperature and is given in the

following table (see also DIN 38404-6 for the system, Ag/AgCl/
saturated KCl):

Contamination

Cleaning

Gross contamination at the
junction

Carefully brush off contamination under
running water using a soft toothbrush

Membrane contamination

Splash the electrode with water, then blot it
dry using a moist paper towel (do not rub).

Fat, oil, protein-containing
coatings and similar substanc-
es

Remove with household washing-up
liquid

with:

U

H

= ORP, referring to the standard hydrogen electrode

U

Meas

= Measured ORP

U

Ref

= Voltage of the reference system compared to the standard

hydrogen electrode

T (°C)

U

Ref

[mV]

T (°C)

U

Ref

[mV]

SensoLyt

®

combina-

tion electrodes

SensoLyt

®

combina-

tion electrodes

0

+221

35

+187

5

+216

40

+181

10

+212

45

+176

15

+207

50

+171

20

+202

55 +165

25

+197

60 +160

30

+192