Technical data

SensoLyt

®

combination electrodes

6

ba76005e01

01/2012

Technical data

pH measuring

range

SensoLyt

®

SEA

SensoLyt

®

ECA

pH 2 ... 12

SensoLyt

®

SEA-HP

pH 4 ... 12

SensoLyt

®

DWA

pH 0 ... 14

ORP measuring

range

SensoLyt

®

PtA

-2000 ... 2000 mV
(pH application range pH 4 ... 12)

Pressure range at

temperature

(armored

electrodes only)

SensoLyt

®

SEA

SensoLyt

®

ECA

SensoLyt

®

DWA

SensoLyt

®

PtA

Temperature

Allowed overpressure

0 °C (32 °F)

1000 kPa (10 bar)

20 °C (68 °F)

1000 kPa (10 bar)

30 °C (86 °F)

500 kPa (5 bar)

40 °C (104 °F)

300 kPa (3 bar)

60 °C (140 °F)

100 kPa (1 bar)

SensoLyt

®

SEA-HP

0-60 °C (32-140 °F)

1000 kPa (10 bar)

All SensoLyt

®

combination electrodes meet the requirements of article 3 (3)

of the directive 97/23/EC ("Pressure equipment directive").

Measuring

electrode

SensoLyt

®

PtA

Platinum ring

All except for
SensoLyt

®

PtA

Glass

Reference

electrolyte,

junction

SensoLyt

®

SEA

SensoLyt

®

SEA-HP

SensoLyt

®

PtA

Gel polymer solid electrolyte;
2-hole junction

SensoLyt

®

ECA

Gel electrolyte; 1-hole junction

SensoLyt

®

DWA

Modified gel electrolyte;
ceramic junction

Shunt conduction

element

Ag/AgCl

Connection

Plug-in system (S7)

Minimum

immersion depth

25 mm