Wear parts and accessories

Oxidizing pretreat-
ment

Immerse the platinum electrode for two to three
days in a sulfuric acid clorina solution (0,5 g clorina
powder, 100 ml H

2

O dist., 2-3 ml 20% sulfuric acid).

Clorina powder for producing the solution is includ-
ed in the SORT/RH reagent set.

Note: The diaphragm must not be immersed in the
clorina solution!

Reducing pretreat-
ment

When the electrode is ready for the test immerse it
in the Zobell ORP buffer solution and wait for a sta-
ble measured value.

Description

Model

Order no.

Reference electrolyte solution 3 mol/l KCl,
Ag

+

free (250 ml)

KCl-250

109 705Y

Reagent set for regenerating ORP platinum
electrodes, consisting of 10 g activation pow-
der and 30 g clorina powder

SORT/RH

109 730Y

Zobell ORP buffer solution (125 ml)

3682

061320