Technical data

4120 / 4130

4

ba76136e01

02/2013

Technical data

General data

Measurement and application characteristics

Shaft dimensions, shaft material, electrical connection

Connection cable

Accuracy of the IDS measuring technique

Model

Reference electrolyte

Junction

NTC

Special features

4120

KCl 3 mol/l, Ag

+

free

Ceramic

Yes

Plastic shaft

4130

KCl 3 mol/l, Ag

+

free

Platinum

Yes

Model

pH measuring range

Allowed temperature
range

4120

0.000 ... 14.000

0 ... 80 °C (32 ... 176 °F)

4130

0.000 ... 14.000

0 ... 100 °C
(32 ... 212 °F)

Model

Membrane resistance
at 25 °C

Typical application

4120

< 1 GOhm

Field

4130

< 600 MOhm

Laboratory

Model

Shaft

Electrical connection

Length
[mm]

Ø

[mm]

Material

Electrode
connection

Meter con-
nection

Cable
length

4120

120

12

Polyamide

Fixed cable

Digital plug

1.5 m

4130

120

12

Glass

Fixed cable

Digital plug

1.5 m

Diameter

4.3 mm

Smallest allowed bend radius

Fixed installation:

20 mm

Flexible use:

60 mm

Plug type

Socket, 4 pins

Measured parameter

Accuracy (± 1 digit)

pH

± 0.004

U [mV]

± 0.2

T [°C]

± 0.1