Aging, Maintenance and cleaning – YSI MultiLab IDS 4120 User Manual

Page 7

Advertising
background image

4120 / 4130

Aging

7

ba76136e01

02/2013

Note
During longer storing periods, salt sediments may develop on the watering
cap. They do not affect the measuring characteristics and can easily be re-
moved with deionized water when the electrode is put into operation again.

Aging

Every pH electrode undergoes a natural aging process. With aging, the re-
sponding behavior becomes slower and the electrode slope and asymmetry
change. Moreover, extreme operating conditions can considerably shorten
the lifetime of the electrode. These are:

 Strong acids or lyes, hydrofluoric acid, organic solvents, oils, fats, bro-

mides, sulfides, iodides, proteins

 High temperatures
 High changes in pH and temperature.

The warranty does not cover failure caused by measuring conditions and
mechanical damage.

Maintenance and cleaning

Refilling the

reference

electrolyte

During operation, a small amount of reference electrolyte leaks through the
junction from the electrode into the test sample. If the level of reference elec-
trolyte becomes too low with time, refill it through the refilling opening. Refill-
ing is very easy with the enclosed dropping bottle. Proceed as follows:

 Cut off the tip of the dropping bottle at a right angle until the opening in the

tip can be seen

 Open the refilling opening of the electrode
 Press the tip of the dropping bottle into the refilling opening while turning

it slightly

 Pump the reference electrolyte in the shaft using the dropping bottle
 Pull the dropping bottle out of the refilling opening while turning it slightly

as necessary.