1 overview, 1 structure and function, 2 recommended fields of application – YSI MultiLab IDS 4320 Conductivity Probe User Manual
Page 3: 1overview
4320
Overview
3
ba76140e01
02/2013
1
Overview
1.1
Structure and function
Structure
Automatic sensor
recognition
The sensor electronics with the stored sensor data is in the connecting head.
The data include, among other things, the sensor type and series number.
With each calibration, the calibration data is written in the sensor and the
calibration history is recorded. The data is recalled by the meter when the
sensor is connected and is used for measurement and for measured value
documentation. Storing the calibration data in the sensor ensures that the
correct cell constant is automatically used if the sensor is operated with
several meters.
The digital transmission technique guarantees the failure-free
communication with the meter even with long connection cables. If the sensor
firmware is enhanced by YSI, it can be updated via the meter.
1.2
Recommended fields of application
Measurements in ultrapure water.
1
Shaft
2
Connecting head with active electronics
2
1