Orp calibration – YSI Pro1030 User Manual

Page 12

Advertising
background image

16

17

7. ‘Calibration Successful’ will display for a few seconds to indicate

a successful calibration and then the instrument will return to
the Run screen.

8. If the calibration is unsuccessful, an error message will display

on the screen. Press the Cal key to exit the calibration error
message and return to the Run screen. See the Troubleshooting
guide for possible solutions.

2-PoINT CALIBRATIoN

1. Place the sensor in pH buffer 7 or 6.86 and allow the temperature

and pH readings to stabilize.

2. Press and hold Cal

for three seconds.

3. Highlight pH and press enter. If pH is not listed as an option,

check the System Setup menu to ensure pH is enabled in the
ISE Sensor Type menu.

4. Highlight 2 point and press enter.
5. If necessary, use the up and down arrow keys to adjust the pH

buffer value. Note the pH mV reading which ideally should be
between -50 and +50 in buffer 7.

6. Press enter to continue to second point.
7. Rinse the sensor and place it in the second pH buffer (4/4.01 or

10/9.18).

8. If necessary, use the up and down arrow keys to adjust the pH

buffer value.

9. Wait approximately 30 to 60 seconds for the pH sensor to

stabilize and for the temperature reading to stabilize. Note the
pH mV reading. pH mVs in buffer 4 should be +159 to 180 mV
from the previous buffer 7 pH mV value. pH mVs in buffer 10
should be -159 to 180 mV from the previous buffer 7 pH mV
value.

10. Press enter to complete the calibration or press Cal

to cancel.

11. ‘Calibration Successful’ will display for a few seconds to indicate

a successful calibration and then the instrument will return to
the Run screen.

12. If the calibration is unsuccessful, an error message will display

on the screen. Press the Cal key to exit the calibration error
message and return to the Run screen. See the Troubleshooting
section of this manual for possible solutions.

3-PoINT CALIBRATIoN

1. Place the sensor in pH buffer 7 or 6.86 and allow the temperature

and pH readings to stabilize.

2. Press and hold Cal

for three seconds.

3. Highlight pH and press enter. If pH is not listed as an option,

check the System Setup menu to ensure pH is enabled in the
ISE Sensor Type menu.

4. Highlight 3 point and press enter.
5. If necessary, use the up and down arrow keys to adjust the pH

buffer value. Note the pH mV reading which should be between
-50 and +50 in buffer 7.

6. Press enter to continue to second point.
7. Rinse the sensor and place it in the second pH buffer (4/4.01 or

10/9.18). If necessary, use the up and down arrow keys to adjust
the pH buffer value.

8. Wait approximately 30 to 60 seconds for the pH sensor to

stabilize and for the temperature reading to stabilize. Note the
pH mV reading. pH mVs in buffer 4 should be +159 to 180 mV
from the previous buffer 7 pH mV value. pH mVs in buffer 10
should be -159 to 180 mV from the previous buffer 7 pH mV
value.

9. Rinse the sensor and place it in the third pH buffer (4/4.01 or

10/9.18). If necessary, use the up and down arrow keys to adjust
the pH buffer value.

10. Wait approximately 30 to 60 seconds for the pH sensor to

stabilize and for the temperature reading to stabilize. Note the
pH mV reading. pH mVs in buffer 4 should be +159 to 180 mV
from the previous buffer 7 pH mV value. pH mVs in buffer 10
should be -159 to 180 mV from the previous buffer 7 pH mV
value.

11. Press enter to complete the calibration or press Cal

to cancel.

12. ‘Calibration Successful’ will display for a few seconds to indicate

a successful calibration and then the instrument will return to
the Run screen.

13. If the calibration is unsuccessful, an error message will display

on the screen. Press the Cal key to exit the calibration error
message and return to the Run screen. See the Troubleshooting
section of this manual for possible solutions.

orp calIBratIon

1. Place the clean sensor in oRP calibration solution. Wait for the

oRP and temperature readings to stabilize.

2. Press and hold Cal

for three seconds.

3. Highlight oRP and press enter. If oRP is not listed as an option,

check the System Setup menu to ensure oRP is enabled in the
ISE Sensor Type menu.

4. Use the up and down arrow keys to adjust the oRP calibration

solution value.

Advertising