Conductivity calibration – YSI Pro2030 User Manual

Page 15

Advertising
background image

22

23

CALIBRATING IN MG/L oR PPM

Power the instrument on and place the sensor into a sample that has

1.

been titrated to determine the dissolved oxygen concentration. The

sample should cover the two holes of the conductivity sensor that are

located near the cable, see figure 8. During the calibration, continuously

stir or move the probe through the sample at a rate of at least 6 inches

(15.5 cm) per second if using a yellow membrane or at least 3 inches

(7.7 cm) per second if using a blue membrane. A stir plate may be

helpful for this calibration procedure.

Allow the dissolved oxygen and temperature readings to stabilize. This

2.

may take 5 to 15 minutes, depending on the type and condition of the

sensor.

Press the Cal key. Highlight Dissolved oxygen and press enter.

3.

Highlight mg/L or ppm depending on what is enabled in the System

4.

Setup menu and press enter.

Use the up and down arrow keys to adjust the mg/L (ppm) reading to

5.

the value of the titrated sample. Press enter to confirm the value and

complete the calibration or press the Cal key to cancel the calibration.

‘Calibration Successful’ will display for a few seconds to indicate a

6.

successful calibration and then the instrument will return to the Run

screen.

If the calibration is unsuccessful, an error message will display on the

7.

screen. Press the Cal key to exit the calibration error message and

return to the Run screen. See the Troubleshooting guide for possible

solutions.

conductIvIty calIBratIon

Ensure the conductivity sensor is clean and dry before performing a conductivity,

specific conductance or salinity calibration.

CALIBRATING

SPECIFIC

(SP.)

CoNDUCTANCE

oR

CoNDUCTIVITY

Note: When calibrating Specific Conductance, the Pro2030 uses the factory

default values for the Specific Conductance Reference Temperature and the

Specific Conductance Temperature Coefficient regardless of what is configured in

the System Setup Menu. The default value for the Reference Temperature is 25°C

It is not necessary to calibrate conductivity, specific conductance

and salinity. Calibrating one of these parameters will simultaneously

calibrate the others. YSI recommends calibrating specific conductance

for greatest ease.

i

and the default value for the Temperature Coefficient is 1.91%/°C. It is important

to note that the Temperature Coefficient of a calibration solution is dependent

on the contents of the solution. Therefore, YSI recommends using a traceable

calibration solution made of KCl (potassium chloride) when calibrating Specific

Conductance since these solutions typically have a Temperature Coefficient of

1.91%/°C. Additionally, be sure to enter the value of the solution as it is listed for

25°C when calibrating Specific Conductance.

Fill a clean container (i.e. plastic cup or glass beaker) with fresh, traceable

1.

conductivity calibration solution and place the sensor into the solution.

The solution must cover the holes of the conductivity sensor that are

closest to the cable (figure 8). Ensure the entire conductivity sensor is

submerged in the solution or the instrument will read approximately

half the expected value. Gently move the probe up and down to remove

any air bubbles from the conductivity sensor.

Figure 8, solution above two holes near cable.

Turn the instrument on and allow the conductivity and temperature

2.

readings to stabilize. Press the Cal key. Highlight Conductivity and press

enter. Next, highlight the desired calibration method, Sp. Conductance

or Conductivity, and press enter.

Highlight the units you wish to calibrate, either uS/cm or mS/cm, and

3.

press enter. 1 mS = 1,000 uS. Next, use the up or down arrow key to

adjust the value on the display to match the value of the conductivity

calibration solution. If calibrating conductivity, it is necessary to look

Advertising