Sensor storage – YSI Pro2030 User Manual

Page 21

Advertising
background image

34

35

Chemical cleaning: Remove the cap membrane and rinse the sensor with

deionized or distilled water. Soak the sensing section of the sensor in a 14%

ammonium hydroxide solution for 2 to 3 minutes or in a 3% ammonia solution

overnight for 8-12 hours (most household ammonia cleaners are typically

around 3%). Rinse heavily in cool tap water followed by a thorough rinsing with

distilled or deionized water. The anode should then be thoroughly wiped with a

wet paper towel to remove the residual layer from the anode. Trapping residual

ammonia under the new membrane cap can quickly tarnish the electrode and/

or give false readings and should therefore be avoided.

After performing a chemical cleaning on the polarographic sensor, lightly sand

the cathode and anode following the mechanical cleaning procedures described

previously.

If this procedure is unsuccessful, as indicated by improper sensor performance,

contact YSI Technical Support or the YSI authorized dealer where you purchased

the instrument.

Galvanic Sensors – Model # 605202
YSI recommends that the Sodium Chloride (NaCl) electrolyte solution and

cap membrane be changed every 2-4 weeks during regular use. In addition, the

electrolyte solution and membrane should be changed if (a) bubbles are visible

under the membrane; (b) significant deposits of dried electrolyte are visible

around the membrane; and (c) if the sensor shows unstable readings or other

sensor-related symptoms.

The Galvanic dissolved oxygen sensor is continuously reducing oxygen even

when the instrument is turned off. This factor allows the sensor to be used with

no warm-up period as soon as the instrument is powered on. However, because

the sensor is “on” all the time, some solid from the oxidation of the zinc anode

will form in the electrolyte within 1-2 weeks of activation. Small amounts of

the solid will generally cause no performance problems, but excessive amounts

may result in jumpy dissolved oxygen readings. The rate of solid formation is

dependent on the type of membrane installed. The formation of solids typically

form more rapidly with 5913 (1.25 mil PE) membrane, and less rapid with 5914

(2 mil PE) membranes.

Chemical cleaning should be performed as infrequently as possible.

First attempt a membrane change and recalibrate. If a new membrane

does not resolve the problem, then proceed with cleaning.

i

At the time the cap membrane is changed, YSI recommends that you rinse the

anode and cathode (figure 12) with distilled or deionized water and wipe with

a clean paper towel. If white deposits are evident on the anode after cleaning,

YSI recommends that you remove this material by sanding the anode with

the 400 grit wet/dry sanding disc included in the membrane kit following the

“Mechanical Cleaning” instructions under the Polarographic Silver Anode

maintenance section of this manual. If there are deposits on the cathode, sand

the cathode with the 400 grit wet/dry sanding disc following the maintenance

instructions listed in this manual for the Polarographic Gold Cathode.

If this procedure is unsuccessful, as indicated by improper sensor performance,

contact YSI Technical Support or the YSI authorized dealer where you purchased

the instrument.

sensor storage
SHoRT TERM SToRAGE

The instrument is supplied with a grey calibration/storage sleeve that slides over

the probe guard. The sleeve is used for short-term storage (less than 30 days). Be

sure to keep a small amount of moisture (clean tap water) on the sponge in the

sleeve during storage. This is simply done to maintain a 100% water saturated

air environment which is ideal for short-term sensor storage. The sensors should

not be submerged in water.

LoNG TERM SToRAGE

The dissolved oxygen and conductivity sensors should be stored long term in a

dry state. When storing for more then 30 days, remove the cap membrane and

thoroughly rinse the dissolved oxygen sensor with distilled or deionized water.

once the sensor has been rinsed either blow it dry with compressed air or allow

The Galvanic electrolyte solution will appear milky white after use but

this will not affect the accuracy of the sensor unless there is excessive

build up. The color change is acceptable and normal as long as DO

readings remain stable.

i

IMPORTANT: Be sure to: (1) Use only the fine sanding discs provided

and (2) Sand as mentioned in the above procedures. Not adhering to

either of these instructions can damage the electrodes.

WARNING: Do not perform the Polarographic chemical cleaning on

a Galvanic sensor.

i

Advertising