Run screen, Backlight, Powering off – YSI Pro2030 User Manual

Page 7: Navigation, Run screen backlight powering off navigation

Advertising
background image

6

7

light blue label on the bottle and Polarographic sensors utilize electrolyte with

a white label on the bottle. The dissolved oxygen sensor is supplied with cap

membranes specific to the sensor type ordered (Polarographic or Galvanic).

5913 and 5914 membrane kits are for Galvanic sensors and the 5908 and 5909

membrane kits are for Polarographic sensors.

Remove and discard or save the red protective cap from the dissolved oxygen

sensor by pulling it straight off. Thoroughly rinse the sensor tip with distilled

or deionized water. Fill the cap membrane 3⁄4 full of electrolyte solution, then

tap the cap with a finger to release any trapped air. Be careful not to touch

the membrane portion of the cap. Thread the membrane cap onto the sensor,

moderately tight. Do not use a tool. It’s typical for some of the electrolyte solution

to spill over. It is best to allow the new cap to remain on the sensor overnight

before calibrating. For detailed instructions on changing a cap membrane, see

the

Care, Maintenance and Storage section of this manual.

run screen

Press the power/backlight key

to turn the instrument on. The instrument

will run through a self test and briefly display a splash screen with system

information before displaying the main Run screen (figure 5). The first time

the Pro2030 is turned on, it will step through language, sensor and membrane

selections; see the First Power on section of this manual for more information.

Figure 5, Run screen.

BacKlIght

once the instrument is powered on, pressing the power/backlight key

will turn on the display backlight. The backlight will remain on until the key is

pressed again or after two minutes of not pressing any key on the keypad.

PowerIng off

To turn the instrument off, press and hold the power/backlight key

for

three seconds.

navIgatIon

The up and down arrow keys allow you to navigate through the functions

of the Pro2030.

NAVIGATING THE RUN SCREEN

When in the Run screen, the up and down arrow keys will move the

highlighted box along the bottom options. once a box is highlighted, press enter

to access the highlighted option.

Description of Run screen box functions from left to right:

Option

Description

Barometer reading

Highlight and press enter to calibrate the barometer.

SAVE

Highlight and press enter to save displayed data to

memory.

DATA

Highlight and press enter to view and/or erase saved

data.

NAVIGATING THE SYSTEM SETUP MENU

When in the System Setup menu, the up and down arrow keys will move the

highlighted bar up and down the system setup options. See the System Setup

menu section of this manual for more information about these options.