fIrst Power on

The instrument will step through an initial configuration when powered on

for the first time. This will set the language, dissolved oxygen sensor type and

membrane type. Use the up or down arrow keys to highlight the appropriate

language, sensor and membrane, then press enter to confirm (figures 6). The

Sensor Type must be configured for the dissolved oxygen sensor installed on the

cable. Failure to do this may result in damage not covered under warranty. If an

incorrect option is selected, it may be changed in the System Setup menu.

Select Language:

English

Français

Español

Deutsch

Use  to select

Language

Press  to confirm

Select Sensor Type:

Polaro (black)

Galvanic (grey)

Use  to select

sensor type

Press  to confirm

Select Membrane

Type:

1.25 (Yellow)

 2.0 (Blue)

Use  to select

membrane

Press  to confirm

Figure 6, Select language, dissolved oxygen sensor and membrane.

After selecting a language, sensor and membrane, the Run screen will be

displayed. The next time the instrument is powered up, the Run screen will

display immediately after the splash screen. If the sensor type or membrane type

is changed, ensure that it is updated in the System Setup menu.

S Y S T E M S E T U P M E N U

Press the menu

key to access the System Setup menu. The System Setup

menu contains multiple screens that are notated as ‘pages’. The current page is

indicated near the bottom of the display (figure 7).

Use the up and down arrow keys to scroll through menu options and menu

pages.

EXITING THE SYSTEM SETUP MENU

To exit the System Setup menu, press the down arrow key until the ESC - Exit

box is highlighted, then press enter to return to the Run screen.

do local%

DO Local% can be enabled or disabled by using the up or down arrow keys to
highlight it and then pressing enter . An ‘X’ in the box next to Do Local%

indicates it is enabled (Figure 7).

Figure 7, page 1 of System Setup menu. DO Local% is enabled.

When Do Local% is enabled, Do% values will be expressed as %L on the Run

screen.

Do Local% allows for localized % saturation measurements, but does not

affect the mg/L readings. When enabled, the Do% calibration value is always

100% regardless of the altitude or barometric pressure of the location. This

deviates from YSI’s traditional method of expressing Do% saturation where

the % calibration value decreases with a decrease in barometric pressure, i.e. an

increase in altitude (Appendix A). To determine the % calibration value when

Do Local% is disabled, divide the local barometric pressure pressure in mmHg

by 760 and then multiply by 100.

Example: 750/760 = 0.9868 x 100 = 98.68% calibration value when Do Local is

disabled.

When Do Local% is enabled, the Pro2030 corrects for barometric pressure on

each Do measurement instead of during calibration.