4 what to do if, 1 no connection to meter, 2 user administration cannot be set up – Xylem HL MKII MultiLab User User Manual

Page 8: 3 forgotten the password, No connection to meter, User administration cannot be set up, Forgotten the password, 4what to do if

Advertising
background image

W

HAT

TO

DO

IF

...

M

ULTI

L

AB

®

U

SER

8

ba76147d01

04/2013

4

What to do if...

4.1

No connection to meter

4.2

User administration cannot be set up

4.3

Forgotten the password?

Cause

Remedy

USB cable defective

Use another USB cable

Baud set on the meter is too high

• Set lower baud rate on the meter

(menu, System/interface)

• Quit the setting menu
• Restart the

MultiLab User

program.

• Re-establish the connection

USB VCP driver is not installed

Install USB VCP driver

Cause

Remedy

The firmware version of the meter

does not support this feature

• Carry out a firmware update (see

operating manual of your meter)

Cause

Remedy

Password for a user name forgotten

• Erase the user name
• Re-enter the user name
• Send the list

The user can assign a new password

with the next login.

Administrator password forgotten

Contact the service department.