Continuation tests (certain tests only), Continuation tests ( certain tests only ) – YSI 9500 Photometer User Manual

Page 13

Advertising
background image


Chlorine-Free / 5

Phot 007

Insert Sample

Blank - Choose a Test - Time

OK

Place

SAMPLE TUBE (sample and reagent) in the test chamber, then

press [OK].

7 The instrument will take the reading and display the result as follows, for

example :-

Chlorine-Free / 5

Phot 007

1.00

mg/l Cl

2

Choose a Test

Read Blank Follow-On

The following symbols indicate the result is out of test range :-

Result is higher than range

> >

Result is lower than range

< <

8 The ‘Options’ menu at the bottom then offers the choice to :-

‘Choose a Test’ - return to the menu of test programs and select another test

‘Read’

- read further sample tubes of the currently selected test

‘Blank’

- re-blank the instrument

‘Follow-On’

- carry out a continuation test if available.


Continuation Tests (Certain Tests Only)

1 Select ‘Follow-On’ and press [OK] during the result display period of the

foregoing test stage. The ‘Insert Sample’ screen will appear.

Place

SAMPLE TUBE in the test chamber, then press [OK].

2 The instrument will take the reading and calculate the result from the

combination of readings (where appropriate). The result will be displayed:

Chlorine-Total / 5

Phot 008

1.50

mg/l Cl

2

Choose a Test Read Blank

Return

13