Physio-Control LIFEPAK 1000 User Manual

Page 56

Advertising
background image

Specifications

A-6

LIFEPAK 1000 Defibrillator Operating Instructions

Waveform Amplitude

1 cm/mV, nominal
Display Range Differential: ±1.4 mV full scale, nominal

Heart Rate

20 to 300 BPM digital display.
Display "---" if heart rate is less than 20 BPM.
Heart symbol flashes for each QRS detection

Displayed ECG

ECG information is received from therapy pads in anterior-lateral
or anterior-posterior positions, or from the 3-wire ECG cable in
Lead II.

Controls

On/Off

Controls device power

Shock

Controls the delivery of defibrillation energy

Soft Keys

Used during device setup and during patient use: Analyze,
Charge, Disarm

Menu button

Used to access additional device features

Readiness Display

The readiness display shows device status

OK Indicator

Indicates OK when the last self-test was completed successfully.

Battery Capacity
Indicator

Segmented display showing battery capacity

Service Indicator

Service required when displayed

Environmental

Note:

All performance specifications defined assume that the device has been stored

(two hours minimum) at the operating temperature prior to operation.

Operating
Temperature

0° to 50°C (32° to 122°F)

One-Hour Operating
Temperature

From room temperature to temperature extreme, one-hour
duration:
-20° to 60°C (-4° to 140°F)

Storage Temperature

With nonrechargeable (Li/MnO

2

) battery and electrodes, maximum

exposure time limited to seven days: -30° to 60°C (-22° to 140°F)

Atmospheric
Pressure

575 hPa to 1060 hPa, 4572 to -382 meters (15,000 feet to -1250
feet)

Relative Humidity

5% to 95% (noncondensing)

Dust/Water

IEC 60529 IP55 with battery and REDI-PAK electrodes installed

Shock

MIL-STD-810F, Method 516.5, Procedure 1, (40g peak, 15–23
msec pulse, 45 Hz crossover frequency)

Bump

EN 1789 and IEC 60068-2-29, Test Eb: (1000 bumps, 15g, 6 ms,
vertical direction)

Advertising