Gorman-Rupp Pumps 82E20-B 717487 and up User Manual

Page 27

Advertising
background image

OM−00822

80 SERIES

MAINTENANCE & REPAIR

PAGE E − 9

This seal is not designed for operation at
temperatures above 160

_F (71_C). Do not

use at higher operating temperatures.

Position the seal plate (26) on a flat surface with the
impeller side facing down. Position the lip seal (7)
in the seal plate bore with the lip position as shown
in Figure 1. Press the lip seal into the seal plate
bore until fully seated.

Position the seal plate on a flat surface with the im-
peller side facing up. Lubricate the stationary seat
O-ring with water or light oil, and press the station-
ary subassembly into the seal plate until fully
seated. After installation, wipe the face of the sta-
tionary element lightly in a concentric pattern with a
clean, lint-free tissue.

Slide the assembled seal plate and stationary seat
over the impeller shaft (20) and temporarily secure
the seal plate to the pedestal (12) using two cap-
screws and nuts (1/2 UNC X 1 1/2 inch long, not
supplied).

Lubricate the O.D. of the shaft sleeve with water or
a very small amount of oil, and slide the rotating
portion of the seal assembly (consisting of the ro-
tating element, bellows and retainer) onto the shaft
sleeve until the face of the rotating element is just
flush with the chamfered end of the sleeve.

Slide the assembled seal and sleeve onto the shaft
until the seal faces contact. Continue to press the
shaft sleeve through the rotating portion of the seal
until it is fully seated against the shaft shoulder. Use
caution not to roll or damage the lip of the seal (7).

Install the seal spring and spring retainer (51).

Lubricate the seal as indicated in LUBRICATION
at the end of this section.

Impeller Installation and Adjustment

Inspect the impeller, and replace it if cracked or
badly worn. Install the same thickness of impeller
shims (52) as previously removed, and screw the
impeller onto the shaft until tight.

A clearance of .010 to .020 inch (0,25 to 0,51 mm)
between the impeller and the seal plate is neces-

sary for maximum pump efficiency. Measure this
clearance and add or remove impeller shims until
this clearance is reached.

NOTE

Be sure the seal plate is tight against the pedestal
while measuring this clearance.

After the back clearance is set, secure the impeller
with the hardware (31, 32 and 33). Reinstall the
bottle oiler and piping (4, 5 and 6).

Pump Casing Installation

If the wear plate assembly (30) was removed, posi-
tion the replacement wear plate assembly squarely
against the casing shoulder and secure it with the
mounting hardware (34, 35, 37 and 38). Replace
the fiber washer (38) if badly worn or compressed.

Remove the hardware temporarily securing the
seal plate to the pedestal. Install the same thick-
ness of pump casing gaskets (29) as previously re-
moved. Secure the pump casing (1) to the seal
plate and pedestal with the nuts (28). Do not fully
tighten the nuts at this time.

A clearance of .008 to .015 inch (0,20 to 0,38 mm)
between the impeller and the wear plate is also rec-
ommended for maximum pump efficiency. Set this
clearance by adding or removing gaskets in the
pump casing gasket set (29) until the impeller
scrapes against the wear plate when the shaft is
turned by hand. After the impeller scrapes, add ap-
proximately .008 inch (0,20 mm) of gaskets.

Secure the pump casing to the base with the pre-
viously removed hardware. Be sure to reinstall any
leveling shims used under the mounting feet of the
pump casing.

If a lifting eye was used to move the pump casing,
be sure to remove the lifting eye from the pump
casing.

Do not attempt to lift the complete pump
unit using the lifting eye. It is designed
to facilitate removal or installation of in-
dividual components only. Additional

Advertising