Lightwave measurement uncertainty is presented in the
following graphs and tables. This covers three types of
measurements: optical (transmission and reflection)
measurements, optical-to-electrical measurements, and
electrical-to-optical measurements. Data is recorded after
an 8703A accuracy enhancement has been performed
using the indicated calibration type. This analysis accounts
for the following errors
23
:
• Residual systematic errors (Table 5)
• System dynamic accuracy (dB from reference)
24
• 3.5 mm connector repeatability
25
• Lightwave source stability
• Lightwave source and receiver factory calibration
uncertainty
26
• Switch repeatability
• Noise
A 10 Hz IF bandwidth, a 16 averaging factor and a
23 ±3°C temperature range are used in all cases. Data
applies to all 8703A internal source configurations of
Table 4.
The following table shows 8703A residual systematic
errors after accuracy enhancement using the same
calibration and setup as stated for each of the three
lightwave measurement types.
Table 5. Residual
lightwave measurement
systematic errors.
Frequency range (GHz)
0.13 to 12.0
12.0 to 20
Optical residual characteristics
(typical)
Lightwave source port return loss
15 dBo
15 dBo
Lightwave receiver port return loss
20 dBo
20 dBo
Transmission tracking
27
±0.55 dB
±0.55 dB
Lightwave directional coupler
test port return loss
37 dBo
37 dBo
Directivity
37 dB
37 dB
Reflection tracking
27
±0.45 dB
±0.45 dB
Electrical residual characteristics
(typical)
Microwave source port return loss
29 dBe
28
29 dBe
Microwave receiver port return loss
30 dBe
30 dBe
8
Lightwave measurement
accuracy summary
Optical transmission and reflection
Measurement setup
Calibration type: response & isolation
Calibration standards:
14.5 dB return loss Fresnel standard
Connectors and cables:
HMS-10 lightwave connectors
40 cm single mode fiber cables
Measurement uncertainty
29
Transmission test
Magnitude
Phase
(typical)
Reflection test
Magnitude
(typical)