NFA Series Key Specifications
Specifications apply over 0°C to +55°C unless otherwise noted. The
analyzer will meet its specifications after 2 hours of storage within
the operating temperature range, 60 minutes after the analyzer is
turned on, with Alignment running. A user calibration is required
before corrected measurements can be made.
Frequency Range
NFA Series:
N8972A
10 MHz to 1.5 GHz
N8973A
10 MHz to 3 GHz
N8974A
10 MHz to 6.7 GHz
N8975A
10 MHz to 26.5 GHz
Measurement Speed (nominal)
8 Averages
64 Averages
N8972A:
<100 ms/measurement
<80 ms/measurement
N8973A:
<50 ms/measurement
<42 ms/measurement
N8974A:
<70 ms/measurement
<50 ms/measurement
N8975A:
<70 ms/measurement
<50 ms/measurement
Measurement Bandwidth (nominal)
N8972A: 4 MHz
N8973A, N8974A, N8975A: 4 MHz, 2 MHz, 1 MHz, 400 kHz, 200 kHz, 100 kHz
Noise Figure and Gain
(Performance is dependent upon ENR of noise source used)
N8972A
Noise Source ENR
4 – 7 dB
12 – 17 dB
20 – 22 dB
Noise Figure
Measurement range
0 to 20 dB
0 to 30 dB
0 to 35 dB
Instrument uncertainty
±<0.1 dB
±<0.1 dB
±<0.15 dB
Gain
Measurement range
–20 to +40 dB
Instrument uncertainty
±<0.17 dB
N8973A, N8974A and
Noise Source ENR
N8975A
(10 MHz to 3.0 GHz)
4 – 7 dB
12 – 17 dB
20 – 22 dB
Noise Figure
Measurement range
0 to 20 dB
0 to 30 dB
0 to 35 dB
Instrument uncertainty
±<0.05 dB
±<0.05 dB
±<0.1 dB
Gain
Measurement range
–20 to +40 dB
Instrument uncertainty
±<0.17 dB
N8974A and N8975A
Noise Source ENR
(>3.0 GHz)
4 – 7 dB
12 – 17 dB
20 – 22 dB
Noise Figure
Measurement range
0 to 20 dB
0 to 30 dB
0 to 35 dB
Instrument uncertainty
±<0.15 dB
±<0.15 dB
±<0.2 dB
Gain
Measurement range
–20 to +40 dB
Instrument uncertainty
±<0.17 dB
Characteristic
1
Noise figure at 23ºC ± 3ºC (10 MHz to 3.0 GHz)
Characteristic
1
Noise figure at 23ºC ± 3ºC (3.0 GHz to 26.5 GHz)
Characteristic values are met or bettered by 90% of instruments with 90%
confidence.
Frequency Reference
Standard
Opt.1D5
Aging
±<2 ppm
1
/year
±<0.1 ppm/year
Temperature stability
±<6 ppm
±<0.01 ppm
Settability
±<0.5 ppm
±<0.01 ppm
Tuning Accuracy (Start, Stop, Center, Marker)
4 MHz Measurement Bandwidth (default on all models of Noise Figure
Analyzer)
Frequency
Error
10 MHz – 3.0 GHz
±<Reference error + 100 kHz
3.0 GHz – 26.5 GHz
±<Reference error + 400 kHz
<4MHz Measurement Bandwidth (functionality not present in N8972A)
Frequency
Error
10 MHz – 3.0 GHz
±<Reference error + 20 kHz
3.0 GHz – 26.5 GHz
±<Reference error + 20% of measurement
bandwidth
1
Parts Per Million (10e-6)
10
8
6
4
2
0
10
500
1000
1500
2000
2500
3000
Frequency (MHz)
Noise Figure (dB)
89
10
11
12
Frequency (MHz)
Noise Figure (dB)
3000
3883
4707
5651
7418
8303
8187
9187
10072
10956
11810
12724
13608
14492
15376
16260
17145
18029
18913
18797
10681
21565
22419
23334
24210
25102
25986
26500