Dc bias settling time – Atec Agilent-E4980A-AL User Manual
Page 36

36
Table 66. Yo_DCI1 value
DC bias current range
Measurement time mode
SHORT
MED,
LONG
20 µA
0 S
0 S
200 µA
0.25 nS
0.05 nS
2 mA
2.5 nS
0.5 nS
20 mA
25 nS
5 nS
100 mA
250 nS
50 nS
Table 67. Yo_DCI2 value
DC bias
Measurement time mode
current range
≤ 100 Ω
300 Ω, 1 k Ω
3 k Ω, 10 k Ω
30 k Ω, 100 k Ω
20 µA
0 S
0 S
0 S
0 S
200 µA
0 S
0 S
0 S
0 S
2 mA
0 S
0 S
0 S
3 nS
20 mA
0 S
0 S
30 nS
30 nS
100 mA
0 S
300 nS
300 nS
300 nS
DC bias settling time
When DC bias is set to ON, add the following value to the settling time:
Table 68. DC bias settling time
Bias
Settling
time
1
Standard
Capacitance of DUT Ч 100 Ч log
e
(2/1.8 m) + 3 m
2
Option 001
Capacitance of DUT Ч 100 Ч log
e
(40/1.8 m) + 3 m
Figure 12. DC bias settling time
10 msec
100 msec
1 sec
10 sec
100 sec
1.
2.
1 µF
10 µF
100 µF
1 mF
10 mF
100 mF
DUT capacitance
Settling time