Features, Applications, Specifications – Atec Aemc-5600 User Manual

Page 2

Advertising
background image

Digital Micro-Ohmmeter Model 5600 (pdf)

03/03

Technical Assistance (800) 343-1391

www.aemc.com

2 of 5

Specifications

ELECTRICAL

Range

2000µ

Ω

20m

Ω

200m

Ω

2000m

Ω

20

Ω

200

Ω

Display

1999

19.99

199.9

1999

19.99

199.9

Resolution

1µ

Ω

10µ

Ω

100µ

Ω

1m

Ω

10m

Ω

100m

Ω

Test Current

10A

1A

1A

100mA

10mA

10mA

Accuracy

0.25% typical

0.25% typical

0.25% typical

0.25% typical

0.25% typical

0.25% typical

0.5% max ± 2cts 0.5% max ± 1ct 0.5% max ± 1ct 0.5% max ± 1ct 0.5% max ± 1ct 0.5% max ± 1ct

Power Source

Rechargeable 6V, 7Ah NiCD battery; built-in 110V/220V (47 to 450Hz) charger;

automatic shutoff after 4 minutes non-use

Battery Life

200µ

Ω

range: 200 to 300 8-second measurements

20-200m

Ω

ranges: 900 30-second measurements (approx.)

2000m

Ω

, 20-200

Ω

ranges: 5000 30-second measurements (approx.)

Operating Voltage

4.5 to 7V

DC

Recharge Time

14 hours from full discharge (approx.)

MECHANICAL

Display

2000-count, 7-segment LED

Connection

Terminals accept 4mm banana plugs or 6mm (min) spade lugs

Operating Temperature

14° to 131°F (-10° to 55°C)

Storage Temperature

-4° to 131°F (-20° to 55°C)

Dimensions

15.35 x 10.24 x 9.84" (390 x 260 x 250mm)

Weight

17 lbs 10 oz (8kg)

Case

Fiberglass-charged polycarbonate with handle and removable lid,

watertight to IP534 (lid closed), IP520 (lid open)

Colors

Safety-yellow case, gray faceplate

SAFETY

Safety Ratings

Electrical: IEC 38, Supply voltage, Electromagnetic: IEC 801-2, Electrostatic discharge;

EN 61010, 300V, Cat. II, IEC 801-3, Electromagnetic fields; IEC 801-5, Electrical shocks,

Mechanical/environmental: IEC 529, Protection Index (IP); IEC 68.2.6, Vibration;

IEC 68.2.27, Shocks; IEC 68-2-29, Shake IEC 68.2.31, Knock over;

IEC 68.2.32, Drop test, Case Material: UL94

Overload Input Protection

380Vrms

Fuses

Two high-interrupting capacity 8A 380V, 1

1

/

4

x

1

/

4"

CE Mark

Yes

Features

•

Four-terminal Kelvin resistance
measurement

•

Up to 10A test current

•

0.25% basic accuracy

•

1m

Ω

resolution

•

Direct reading, easy to operate

•

Six selectable ranges

•

Reverse polarity switch

•

Overload and input fuse protection

•

Automatic decimal point and zero

•

Rechargeable NiCD battery with
integral charger (110/220V)

•

Rugged field case for industrial
applications

Applications

•

Contact resistance of breakers
and switches

•

Wire to terminal connections

•

Winding resistance of motors
and transformers

•

Resistance on electric motors

•

Aircraft and rail bonds

•

Grounding connections

•

Battery strap measurements

Test Equipment Depot www.testequipmentdepot.com