Atec AEMC-8335 User Manual
Page 7

6
PowerPad Model 8335
Accuracy Specifications of the Current Probes (to be added to the meter’s accuracy)
Sensor Type
TRMS Current
Maximum Error on I
RMS
Maximum Error on
MR193
1000A
[1A; 10A]
±(1.5%+1A)
N.S.
[10A; 100A]
±(2°)
[100A; 800A]
±(3%)
±(1.5°)
[800A; 1200A]
±(5%)
[1200A; 1400A]
SR193
1000A
[1A; 3A]
±(0.8%)
N.S.
[3A; 10A]
±(1°)
[10A; 100A]
±(0.3%)
±(0.5°)
[100A; 1200A]
±(0.2%)
±(0.3°)
AmpFlex
®
3000A
[10A; 100A]
±(3%)
±(1°)
[100A; 6500A]
±(2%)
±(0.5°)
MiniFlex
™
1000A
[10A; 100A]
±(3%)
±(1°)
[100A; 1000A]
±(2%)
±(0.5°)
MN93
200A
[0,5A; 2A]
±(3%+1A)
N.S.
[2A; 10A]
±(6°)
[10A; 100A]
±(2.5%+1A)
±(3°)
[100A; 240A]
±(1%+1A)
±(2°)
MN193
100A
[100mA; 300mA]
±(0.7%+2mA)
N.S.
[300mA; 1A]
±(1.5°)
[1A; 120A]
±(0.7%)
±(0.7°
MN193
5A
[5mA; 50mA]
±(1%+0.1mA)
±(1.7°)
[50mA; 500mA]
±(1%)
±(1°)
[500mA; 6A]
±(0.7%)
5A
Adapter
[5mA; 50mA]
±(1%)
±(1°)
[50mA; 6A]
±(0.5%)
±(0°)