Dynamic
Range (dB)
(A)-(B)
Leveled
Output
Power
(dBm)
(A)
Noise Floor
(dBm)
(B)
Dynamic
Range (dB)
(C)-(D)
Compression
at Test Port
(dBm)
(C)
Noise Floor
(dBm)
(D)
10 MHz to 50 MHz
82
12
-70
91
15
-76
50 MHz to 100 MHz
105
13
-92
112
15
-97
100 MHz to 500 MHz
114
13
-101
118
12
-106
500 MHz to 1 GHz
123
13
-110
127
12
-115
1 GHz to 10 GHz
127
13
-114
131
12
-119
10 GHz to 13.5 GHz
126
12
-114
132
12
-120
13.5 GHz to 16 GHz
128
12
-116
133
12
-121
16 GHz to 24 GHz
127
11
-116
133
12
-121
24 GHz to 26.5 GHz
128
11
-117
133
12
-121
26.5 GHz to 30 GHz
116
10
-106
124
12
-112
30 GHz to 32 GHz
115
9
-106
123
11
-112
32 GHz to 35 GHz
116
10
-106
123
11
-112
35 GHz to 40 GHz
109
5
-104
121
11
-110
40 GHz to 50 GHz
112
11
-101
119
11
-108
50 GHz to 60 GHz
112
11
-101
118
11
-107
60 GHz to 67 GHz
112
11
-101
119
11
-108
67 GHz to 70 GHz
--
--
--
117
11
-106