Stability description specification spd, Dynamic accuracy, Description specification (db) typical – Atec Agilent-N5072A Series User Manual
Page 12
Advertising

12
Stability
Description
Specification
SPD
Magnitude
30 to 300 kHz
300 kHz to 3 GHz
3 to 6 GHz
6 to 8.5 GHz
± 0.04 dB/°C
± 0.005 dB/°C
± 0.03 dB/°C
± 0.04 dB/°C
Phase
30 to 300 kHz
300 kHz to 3 GHz
3 to 6 GHz
6 to 8.5 GHz
± 0.8 deg/°C
± 0.1 deg/°C
± 0.4 deg/°C
± 0.8 deg/°C
Dynamic Accuracy
1
Description
Specification (dB)
Typical
Magnitude
10 dBm
–30 dBm
–100 dBm
–110 dBm
± 0.21 dB
± 0.045 dB
± 2 dB
± 3.0 dB
Phase
10 dBm
–30 dBm
–100 dBm
± 5 deg
± 0.3 deg
± 15 deg
1. Accuracy of the test port input power reading is relative to –10 dBm reference input power level.
0.01
0.1
1
10
-100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Accuracy (dB)
Testport Power (dBm)
Magnitude
0.1
1
10
100
-100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
A
ccur
acy
(
d
egr
ees
)
Testport Power (dBm)
Phase
Figure 10. Dynamic Accuracy
Advertising