Measurement throughput summary – Atec Agilent-E5062A User Manual
Page 18

18
Table 1-20
Typical cycle time for measurement completion
1
(ms) (Display update: off)
Number of points
51
201
401
1601
Start 1 GHz, stop 1.2 GHz, 30 kHz IF bandwidth
Uncorrected
8
19
33
117
2-port cal
14
35
63
230
Start 300 kHz, stop 1.5 GHz, 30 kHz IF bandwidth
Uncorrected
15
25
39
123
2-port cal
27
48
75
243
Start 300 kHz, stop 3 GHz, 30 kHz IF bandwidth
Uncorrected
17
28
41
125
2-port cal
31
53
80
247
Measurement throughput summary
1. Typical performance.
Table 1-21
Typical cycle time for measurement completion
1
(ms) (Display update: on)
Number of points
51
201
401
1601
Start 1 GHz, stop 1.2 GHz, 30 kHz IF bandwidth
Uncorrected
59
68
83
172
2-port cal
85
103
131
304
Start 300 kHz, stop 1.5 GHz, 30 kHz IF bandwidth
Uncorrected
64
74
89
178
2-port cal
95
116
144
317
Start 300 kHz, stop 3 GHz, 30 kHz IF bandwidth
Uncorrected
66
78
91
180
2-port cal
98
121
148
322