E-series e9300 average power sensors – Atec Agilent-8480 Series User Manual
Page 14
14
Power Linearity (after zero and cal at ambient environmental conditions)
Table 10
Sensor
Power
Linearity
Linearity
(25 ˚C ± 10 ˚C)
(0 – 55 ˚C)
E9300A, E9301A
-60 to –10 dBm
±3.0%
±3.5%
E9304A
-10 to 0 dBm
±2.5%
±3.0%
0 to +20 dBm
±2.0%
±2.5%
E9300B, E9301B
-30 to +20 dBm
±3.5%
±4.0%
+20 to +30 dBm
±3.0%
±3.5%
+30 to +44 dBm
±2.5%
±3.0%
E9300H, E9301H
-50 to 0 dBm
±4.0%
±5.0%
0 to +10 dBm
±3.5%
±4.0%
+10 to +30 dBm
±3.0%
±3.5%
E-Series E9300 Average Power Sensors
0.5
Power (dBm)
% Error
0.4
0.3
0.2
0.1
0
-0.1
-0.2
-0.3
-0.4
-0.5
-30
-26
-22
-18
-14
-10
-6
-2
2
6
10
14
18
20
Typical E9300A/01A/04A Power Linearity at 25 ˚C, after zero and calibration, with associated
Measurement Uncertainty
Power Range
-30 to -20 dBm
-20 to -10 dBm
-10 to 0 dBm
0 to +10 dBm
+10 to +20 dBm
Measurement
±0.9%
±0.8%
±0.65%
±0.55%
±0.45%
Uncertainty