Power meter specifications continued – Atec Anritsu-MA2445A User Manual

Page 4

Advertising
background image

4

ML2430A Series

ML2480A Series

ML2490A Series

Comments

ML2437A

ML2438A

ML2487A

ML2488A

ML2495A

ML2496A

Operation

2

2 (CW or Pulse/Modulated measurement modes)

Measurement Display-

Readout (Numerical)

Power vs. Time graphic of readout data

or Profile of Peak power for analysis of

repetitive pulse or transient waveforms

2 (Pulse/Modulated measurement mode)

Measurement Display-

Profile (Graph)

Single channel power sweep or frequency sweep

Source sweep

±5 dB range CW (Readout mode) only

Peaking meter

Dynamic range covered by five overlapping

amplifier ranges, R1, R2, R3, R4 and R5
Universal Sensor MA2481/82D ranges 1 to 6

Pulse modulated mode:

Dynamic range covered by three overlapping amplifier ranges, R7, R8 and R9
CW mode:

Dynamic range covered by five overlapping amplifier ranges, R1, R2, R3, R4 and R5
Universal Sensor MA2481/82D ranges 1 to 6

Amplifier Range

Auto or Manual

(current range or selectable 1 through 5)

Automatic or manual. When in manual clear indication given to user

(display and GPIB) of fault conditions (under or over-range)

Range Hold

Features

(summary)

0.1 to 0.001 dB
Linear power units, 3 to 6 digit, 1 to 3 digits

selectable to right of decimal nW to W;
Voltage, 1 to 2 digits selectable to right

of decimal

0.1 to 0.001 dB

Display resolution in

Readout mode

0.01 dB

Display resolution in

Profile mode

Profile and P vs. T modes:

200 pixels display resolution
For a 1 ms Profile window, cursor resolution

on the display is 5 µs

16 ns

Pulse/Modulated mode
15 µs

CW Mode

1 ns (RSS mode)

16 ns (non RSS mode)

Pulse/Modulated mode
15 µs

CW Mode

Time measurement

resolution

Hold, Max, Min

Measurement hold

Average, Min, Max

Average, Min, Max, Peak, Crest, PAE (Power Added Efficiency)

Measurements

—

PDF, CDF, CCDF

Power statistics

0.00 to 20.00V nominal

Voltage

measurement range

Watt, %, Volts

Display units (Lin)

Display units (Log)

dBm, dB, dBµV, dBmV, dBr

dBm, dBW, dB, dBµV, dBmV

–199.99 to +199.99 dB

Display range

1

Four Independently set Gates or eight repeated Gates
One Fence per Measurement gate
Gate measurement supports Average, Peak, Crest, Max and Min

Measurement Gates

2

Four Markers and One Delta Marker, Marker to Max/Min, Pulse Rise/Fall-time,

Pulse Width, Off Period, Pulse Repetition Interval
Rise Fall/Search Parameter Variable %
Reference: Max Marker or Gate Power Level

Markers

Fixed value high and low limits with audible,

rear panel TTL output, and/or visible

Pass/Fail alarm indication
Failure indication can latch for

transient failure detection

Simple pass/fail for CW
Complex limits for pulsed and TDMA systems
30 Limits Stores available on the instrument

Limit lines

–199.99 to +199.99 dB (Fixed value or frequency dependent table)

Offset range

Power Meter Specifications continued