Specifications (continued)

Zero set, zero drift, and measurement noise

Average only Mode

Power range

1

Zero set (internal)

Zero set (external)

Zero drift

2

Measurement noise

3

U2000/1/2/4A

–60 to –35 dBm
–38 to –15 dBm
–20 to –6.5 dBm
–7.5 to –2 dBm
–4 to 15 dBm
10 to 20 dBm

± 1.5 nW
± 2 nW
± 12 nW
± 2 µW
± 4 µW
± 6 µW

± 600 pW
± 1.5 nW
± 10 nW
± 500 nW
± 1 µW
± 5 µW

200 pW
400 pW
1.5 nW
50 nW
500 nW
2 µW

1 nW
1.5 nW
15 nW
650 nW
1 µW
10 µW

U2004A

–60 to –35 dBm
–38 to –15 dBm
–20 to –6.5 dBm
–7.5 to –2 dBm
–4 to 15 dBm
10 to 20 dBm

± 2.8 nW
± 3 nW
± 12 nW
± 2 µW
± 4 µW
± 6 µW

± 600 nW
± 1.5 nW
± 10 nW
± 500 µW
± 1 µW
± 5 µW

200 nW
400 nW
1.5 nW
50 nW
500 µW
2 µW

1 nW
1.5 nW
15 nW
650 µW
1 µW
10 µW

U2000/1/2H

–50 to –25 dBm
–28 to –5 dBm
–10 to 3.5 dBm
2.5 to 8 dBm
6 to 25 dBm
20 to 30 dBm

± 15 nW
± 20 nW
± 120 nW
± 20 µW
± 40 µW
± 60 µW

± 8 nW
± 20 nW
± 100 nW
± 20 µW
± 30 µW
± 60 µW

2 nW
4 nW
15 nW
500 nW
5 µW
20 µW

10 nW
15 nW
150 nW
6.5 µW
10 µW
100 µW

U2000/1B sensors

–30 to –5 dBm
–8 to 15 dBm
10 to 23.5 dBm
22.5 to 28 dBm
26 to 44 dBm

± 1.8 µW
± 2 µW
± 12 µW
± 2 mW
± 4 mW

± 800 nW
± 2 µW
± 10 µW
± 1 mW
± 2 mW

200 nW
400 nW
1.5 µW
50 nW
500 µW

1 µW
1.5 µW
15 µW
650 µW
1 mW

1. Condition: (i) 0 to 55 °C and (ii) 95% relative humidity at 40 °C non-condensing.

2. Within one hour after zero set, at a constant temperature, after a 24-hour warm-up of the power sensor.

3. The number of averages at 1 for Normal speed, gate length of 2.27 ms, measured over one minute interval and two standard deviations.