Trigger, Sampling characteristics, Rear panel inputs/outputs – Atec Agilent-E9320 Series User Manual

Page 4: Remote programming, Environmental specifications, Regulatory information

Advertising
background image

4

Trigger

Sources:

Internal, External TTL,
GPIB, RS232/422,

Time resolution:

50 ns

Delay range:

±1.0 s

Delay resolution:

50 ns for delays < ±50 ms;
otherwise 200 ns

Hold-off:

Range:

1 us to 400 ms

Resolution:

1% of selected value
(minimum of 100 ns)

Internal trigger:

Range:

-20 to +20 dBm

Level accuracy:

±0.5 dB

Resolution:

0.1 dB

Latency:

500 ns ± 100 ns

Latency is defined as the delay between the applied RF
crossing the trigger level and the meter switching into
the triggered state.

External trigger range:

High > 2.0 V, Low < 0.8 V;

BNC connector; rising or falling edge triggered; input
impedance > 1 k

W.

Trigger out:

Output provides TTL compatible levels

(high > 2.4 V, low < 0.4 V) and uses a BNC connector

Sampling characteristics

Sampling rate:

20 Msamples/second

Sampling technique:

Continuous sampling

Rear panel inputs/outputs

Recorder output(s):

Analog 0 to 1 V, 1 k

W output

impedance, BNC connector. Two outputs are available
on E4417A (channels A and B).

Remote input/output:

TTL output:

used to signal when mea-
surement has exceeded a
defined limit.

TTL input:

initiates zero and calibration
cycle.

Connector type:

RJ-45 series shielded modu-
lar jack assembly.

TTL output:

high = 4.8 V max;

low = 0.2 V max.

TTL input:

high = 3.5 V min, 5 V max;

low = 1 V max, -0.3 V min.

RS-232/422 interface:

Serial interface for communi-

cation with an external controller. Male plug 9-pin
D-subminiature connector.

Trigger in:

Accepts a TTL signal for initiating measure-

ments, BNC connector.

Trigger out:

Outputs a TTL signal for synchronizing

with external equipment, BNC connector.

Ground:

Binding post accepts 4 mm plug or bare wire

connection

Line power

Input voltage range

85 to 264 Vac,
automatic selection

Input frequency range

47 to 440 Hz

Power requirement

approximately 50 VA
(14 Watts)

Remote programming

Interface:

GPIB interface operates to IEEE 488.2 and

IEC-625. RS-232 and RS-422 serial interfaces supplied as
standard

Command language:

SCPI standard interface

commands

GPIB compatibility:

SH1, AH1, T6, TE0, L4, LE0, SR1,

RL1, PP1, DC1, DT1, C0.

Environmental specifications

Operating environment

Temperature

0° to 55 °C

Maximum humidity

95% at 40 °C,
(non-condensing)

Minimum humidity

15% at 40 °C

Maximum altitude

3,000 meters
(9,840 feet)

Storage conditions:

Storage temperature

-20 to +70°C

Non-operating maximum
humidity:

90% at 65 °C
(non-condensing)

Non-operating maximum
altitude:

15,420 meters
(50,000 feet)

Regulatory information

Electromagnetic compatibility:

This product

conforms with the protection requirements of European
Council Directive 89/336/EEC for Electromagnetic
Compatibility (EMC). The conformity assessment
requirements have been met using the technical
Construction file route to compliance, using EMC test
specifications EN 55011:1991 (Group 1, Class A) and EN
50082-1:1992. In order to preserve the EMC performance
of the product, any cable which becomes worn or dam-
aged must be replaced with the same type and specifica-
tion.

Product safety:

This product conforms to the require-

ments of European Council Directive 73/23/EEC, and
meets the following safety standards:

IEC 61010-1(1990) + A1 (1992) + A2 (1995) /
EN 61010-1 (1993)
IEC 825-1 (1993) / EN 60825-1 (1994)
Canada / CSA C22.2 No. 1010.1-93