Table 10. Power linearity (after zero and cal at ambient environmental conditions)
Sensor Power
Linearity
Linearity
(25°C ± 10°C)
(0 – 55°C)
E9300A, E9301A, E9304A
-60 to –10 dBm
±3.0%
±3.5%
-10 to 0 dBm
±2.5%
±3.0%
0 to +20 dBm
±2.0%
±2.5%
E9300B, E9301B
-30 to +20 dBm
±3.5%
±4.0%
+20 to +30 dBm
±3.0%
±3.5%
+30 to +44 dBm
±2.5%
±3.0%
E9300H, E9301H
-50 to 0 dBm
±4.0%
±5.0%
0 to +10 dBm
±3.5%
±4.0%
+10 to +30 dBm
±3.0%
±3.5%
Typical E9300A/01A/04A Power Linearity at 25°C, after zero and calibration, with associated
measurement uncertainty
Power range
-30 to -20 dBm
-20 to -10 dBm
-10 to 0 dBm
0 to +10 dBm
+10 to +20 dBm
Measurement uncertainty
±0.9% ±0.8% ±0.65%
±0.55%
±0.45%