DIGITAL TRANSMISSION MEASURING INSTRUMENTS

324

Numerical relation between data length and step width

Data length

Step width

2 to 65536

1 step

65536 to 131072

2 step

131072 to 262144

4 step

262144 to 524288

8 step

524228 to 1048576

16 step

1048576 to 2097152

32 step

2097152 to 4194304

64 step

4194304 to 8388608

128 step

Output waveform

NRZ

Number of outputs

2 (DATA/DATA)

DATA/DATA tracking mode

ON/OFF selectable

Amplitude

0.25 to 2 Vp-p, 2 mV steps (setting error: ±15% or ±100 mV, whichever is greater)

Voltage: –2 to ±2 V (V

OH

), 1 mV steps (setting error: ±15% of offset voltage, ±100 mV or ±15% of amplitude

Data

Offset voltage

whichever is the greatest

output

Display: V

OH

, V

TH

or V

OL

selectable

Rise/fall time

Typical 30 ps (10% to 90% of amplitude)

Pattern jitter

20 psp-p (typical 10 psp-p)

Waveform distortion

15% or

150 mV whichever is greater

Load impedance

50

(with back termination)

Connector

APC-3.5

Number of outputs

3 (CLOCK 1, CLOCK 1, CLOCK 2)

CLOCK delay

±500 ps (1 ps steps)

Amplitude

0.25 to 2 Vp-p (2 mV steps)

(CLOCK 1, CLOCK 1)

Setting error: ±15% (1.5 to 2 Vp-p), ±25% (0.5 to 1.5 Vp-p), ±100 mV (0.25 to 0.5 Vp-p)

Amplitude (CLOCK 2)

1 Vp-p ±35%

Clock

Voltage: –2 to ±2 V (V

OH

), 1 mV steps (setting error: ±15% of offset voltage, ±100 mV or ±15% of amplitude

output

Offset voltage

whichever is the greatest)

(CLOCK 1, CLOCK 1)

Display: V

OH

, V

TH

or V

OL

selectable

Offset voltage (CLOCK 2)

0 V ±200 mV (V

OH

)

Rise/fall time

Typical 30 ps (10% to 90% of amplitude)

Waveform distortion

15% or

150 mV whichever is greater

Duty factor adjust function

CLOCK 1, CLOCK 1 adjustable

Clock

Load impedance

50

(CLOCK 1, CLOCK 1: with back termination)

output

Connector

APC-3.5 (CLOCK 1, CLOCK 1), SMA (CLOCK 2)

Number of outputs

DATA: 4, CLOCK: 1

Output level

0.5 to 2 Vp-p, 2 mV steps (setting error: ±15% or ±100 mV, whichever is greater)

Voltage: –1.5 to +1.5 V (V

OH

), 1 mV steps (setting error: ±15% of offset voltage or ±15% of amplitude or

Offset voltage

±100 mV whichever is the greatest)

1/4 data

Display: V

OH

, V

TH

or V

OL

selectable

and clock

Rise/fall time

150 ps (20% to 80% of amplitude)

output*

2

Data output jitter

100 psp-p

Waveform distortion

15%

Skew
(DATA/DATA, DATA/CLOCK)

100 ps

Connector

SMA

Number of outputs: DATA 8, CLOCK 1

1/8 data, clock output*

3

Output level: ECL
Connector: SMA

Sync. signal

Number of outputs

1 (1/32 clock, fixed position pattern, or variable position pattern selectable)

output

Output level

Amplitude: 1 Vp-p ±20%, offset voltage: 0 V ±200 mV (V

OH

)

External control

GPIB, IEEE 488.2

Operating temperature range

0˚ to +50˚C

Media: 3.5 inch FD (2HD, 2DD)

Parameter memory

Format: MS-DOS (Rev. 3.1)*

4

Content: Programmable pattern and other parameters

Power

*

5

Vac ±10%, 50/60 Hz,

700 VA

Dimensions and mass

426 (W) x 221 (H) x 451 (D) mm,

33 kg

EN55011: 1991, Group 1, Class A

EMC

EN50082-1: 1992
Harmonic current emissions: EN61000-3-2 (1995)

Safety

EN61010-1: 1993 (Installation Category

ΙΙ

, Pollution Degree

ΙΙ

)

*

1: Relationship between number of pages and items of word length, number of words, and data length

Relationship between pages of WORD mode and DATA mode

Output pattern/mode

Variable page range

WORD

1 word to the number of words that have
been set, 1 step width

1 to < data length/16, 1 step width (up to
quotient value when the remainder is 0, up to
quotient value +1, 1 step width)

DATA

Data length Number of pages

2 to 16

1

17 to 32

2

33 to 48

3

49

4

Advertising