6

Modulation types and operating modes

Carrier

AM

FM

PM

FSK

BPSK

PWM

Sum

Burst

Sweep

Sine and Square

●

●

●

●

●

●

●

●

Pulse

●

●

●

●

●

●

●

●

●

Triangle and Ramp

●

●

●

●

●

●

●

●

Gaussian Noise

●

●

●

a

PRBS

●

●

●

●

●

Single ARB

●

●

b

●

b

●

●

Sequenced ARB

●

●

Modulating signals

Carrier

Sine

Square

Triangle / Ramp

Noise

PRBS

ARB

External

Sine

●

●

●

●

●

●

●

Square and Pulse

●

●

●

●

●

●

●

Triangle and Ramp

●

●

●

●

●

●

●

Gaussian Noise

●

●

●

●

●

●

PRBS

●

●

●

●

●

●

ARB

●

●

●

●

●

●

Modulation characteristics

Amplitude modulation (AM)

Source

Internal or external, or either channel with 33522A

Type

Full-carrier or double-sideband suppressed-carrier

Depth

1

0% to 120%, 0.01% resolution

Frequency modulation (FM)

7

Source

Internal or external, or either channel with 33522A

Deviation

1 µHz to 15 MHz, 1 µHz resolution

Phase modulation(PM)

Source

Internal or external, or either channel with 33522A

Deviation

0° to 360°, 0.1° resolution

Frequency shift key modulation (FSK)

7

Source

Internal timer or ext trig connector

Mark & space

Any frequency within the carrier signal's range

Rate

0 Hz to 1 MHz

Binary phase shift key modulation (BPSK)

Source

Internal timer or ext trig connector

Phase shift

0° to 360°, 0.1° resolution

Rate

0 Hz to 1 MHz

Pulse width modulation (PWM)

Source

Internal or external, or either channel with 33522A

Deviation

8

0% to 100% of pulse width, 0.01% resolution

Additive modulation (Sum)

Source

Internal or external, or either channel with 33522A

Ratio

6

0% to 100% of carrier amplitude, 0.01% resolution

a. Gated burst only. b. Applies to sample clock, not whole waveform.