Atec Agilent-N5183A User Manual
Page 9

9
Residual FM (CW mode, 0.3 to 3 kHz bandwidth, CITT, RMS)
< N*5Hz (typ)
Broadband noise
CW mode at +10 dBm or maximum specified output
power, whichever is lower for offsets > 10 MHz
0.25 to 10 GHz
–145 dBc/Hz (typ)
10 to 20 GHz
–135 dBc/Hz (typ)
20 to 40 GHz
–130 dBc/Hz (typ)
Harmonics (dBc)
1
250 kHz to 2 GHz
–28 (–30 typ)
> 2 to 20 GHz
–54 (–60 typ)
> 20 to 40 GHz
–56 (typ)
Non-harmonics (dBc)
1, 2
250 kHz to 250 MHz
–62 (–89 typ)
> 250 to 375 MHz
–68 (–86 typ)
> 375 to 750 MHz
–57 (–74 typ)
> 750 MHz to 1.5 GHz
–54 (–70 typ)
> 1.5 to 3.2 GHz
–54 (–68 typ)
> 3.2 to 6 GHz
–47 (–63 typ)
> 6 to 12 GHz
–41 (–57 typ)
> 12 to 24 GHz
(–50 typ)
> 24 to 40 GHz
(–45 typ)
Subharmonics (dBc)
1
250 kHz to 1.5 GHz
None
> 1.5 GHz to 20 GHz
–53
> 20 to 40 GHz
–50
1.
CW mode at +10 dBm or maximum specified output power, whichever is lower.
2. Non-harmonics specifications apply to units with serial numbers ending with 49060000 or greater.
For units with lower serial numbers, refer to the archive section at the end of this document.