3

Maximum safe input level

Average continuous power

(input attenuator

≥

15 dB)

E4401B

+30 dBm (1 W)

E4401B (75

Ω

Option 1DP)

+75 dBmV (0.4 W)
(input attenuator

≥

5 dB)

E4402B/04B/05B/07B

+30 dBm (1 W)

Peak pulse power

(input attenuator

≥

30 dB)

E4401B

+30 dBm (1 W)

E4401B (75

Ω

Option 1DP)

+75 dBmV (0.4 W)

E4402B/04B/05B/07B

+50 dBm (100 W)

dc

E4401B, E4402B

100 Vdc

E4401B (75

Ω

Opt. 1DP)

100 Vdc

E4402B (Option UKB)

0 Vdc (dc coupled)
50 V (ac coupled)

E4404B, E4405B

0 Vdc (dc coupled)
50 V (ac coupled)

E4407B

0 Vdc

1 dB gain compression

(total power at input mixer

5

)

50 MHz to 6.7 GHz

0 dBm

6.7 GHz to 13.2 GHz

–3 dBm

13.2 GHz to 26.5 GHz

–5 dBm

Displayed Average Noise Level (DANL)

(dBm)

(Input terminated, 0 dB attenuation, sample detector)
1 kHz RBW; 30 Hz VBW
10 Hz RBW; 1 Hz VBW (Option 1DR)

1 kHz RBW 10 Hz RBW 10 Hz RBW

10 Hz RBW

(Option 1DR) (Option 1DR) (Option 1DR)

(w/preamp

(w/preamp

Option 1DS) Option 1DS)

Typical

E4401B
400 kHz to 1 MHz

≤

–115

≤

–134

≤

–150

≤

–155

1 MHz to 500 MHz

≤

–119

≤

–138

≤

–154

≤

–156

500 MHz to 1 GHz

≤

–117

≤

–136

≤

–152

≤

–156

1 GHz to 1.5 GHz

≤

–114

≤

–133

≤

–150

≤

–155

E4402B
30 Hz to 9 kHz

22

na

≤

–93 na

na

(Option UKB)

9 kHz to 100 kHz

22

na

≤

–109 na

na

100 kHz to 1 MHz

22

na

≤

–135 na

na

1 MHz to 10 MHz

22

≤

–117

≤

–136

na

≤

–152

10 MHz to 1 GHz

≤

–117

≤

–136

≤

–152

19

≤

–156

1 GHz to 2 GHz

≤

–116

≤

–135

≤

–153

19

≤

–156

2 GHz to 3 GHz

≤

–114

≤

–133

≤

–151

19

≤

–154

E4404/05B/07B
30 Hz to 9 kHz

22

na

≤

–93 na

na

(Option UKB)

9 kHz to 100 kHz

22

na

≤

–109 na

na

100 kHz to 1 mHz

22

na

≤

–135 na

na

1 MHz to 10 MHz

22

≤

–117

≤

–137

na

≤

–155

10 MHz to 1 GHz

≤

–116

≤

–135

≤

–151

19

≤

–157

1 GHz to 2 GHz

≤

–116

≤

–135

≤

–151

19

≤

–155

2 GHz to 3 GHz

≤

–112

≤

–131

≤

–149

19

≤

–152

3 GHz to 6 GHz

≤

–112

≤

–131

na

≤

–138

6 GHz to 12 GHz

≤

–111

≤

–130

na

≤

–137

12 GHz to 22 GHz

≤

–107

≤

–126

na

≤

–134

22 GHz to 26.5 GHz

≤

–106

≤

–125

na

≤

–132

E4407B (Option AYZ)
External mixer

6

≤

–134 +

≤

–153 +

na

na

external mixer external mixer

conversion conversion
loss loss

Display range

Log scale

0.1, 0.2, 0.5 dB/division and 1 to 20 dB/
division in 1dB steps;
ten divisions displayed.

RBW = 1 kHz

0 to -85 dB from reference level is
calibrated

RBW = 300 Hz (Option 1DR) 0 to -120

13

dB from reference level

is calibrated

Linear scale 10 divisions
Scale units

dBm, dBmV, dBµV, Volts, and Watts

(Option BAA)

add Hz

Marker readout resolution

Log scale

0 to –85 dB

0.04 dB

0 to –120 dB (Option 1DR)

0.04 dB

Linear scale

0.01% of reference level

Fast sweep times for zero span (Option AYX)
(sweeptimes

≤

sweep points –1/100 kHz)

Log scale

0 to –85 dB

0.3 dB

Linear

0.3% of reference level

Frequency response

(10 dB input attenuation)

Absolute

7

/Typical

Relative flatness

8

E4401B

9 kHz to 1.5 GHz

±0.5 dB

na

±0.5 dB

E4402B/04B/05B/07B

30 Hz to 3 GHz

6

±0.5 dB

na

±0.5 dB

(Option UKB)

9 kHz to 3.0 GHz

±0.46 dB ±0.14 dB

±0.5 dB

3.0 GHz to 6.7 GHz

±1.5 dB

±0.38 dB

±1.3 dB

6.7 GHz to 13.2 GHz

±2.0 dB

±0.68 dB

±1.8 dB

13.2 GHz to 26.5 GHz ±2.0 dB

±0.86 dB

±1.8 dB

Input attenuation switching uncertainty at 50 MHz

Attenuation setting

0 dB to 5 dB

±0.3 dB

10 dB

reference

15 dB

±0.3 dB

20 to 60 dB (E4401B)

±(0.1 dB + 0.01 x attenuator setting)

20 to 65 dB

±(0.1 dB + 0.01 x attenuator setting)

Absolute amplitude accuracy

Typical

At reference settings

15

±0.34 dB

±0.13 dB

E4401B

±0.30 dB

±0.10 dB

Preamp on

16

(Option 1DS)

±0.37 dB

±0.14 dB

External mixer (Option AYZ)

IF INPUT absolute amplitude
accuracy + external mixer
conversion loss accuracy

17

Overall amplitude accuracy

9

±(0.54 dB + absolute frequency

response)

Advertising