System Performance with the E5053A Downconverter
Table 2-2. System phase noise sensitivity (typical)
Offset from carrier (Hz)
Input frequency
1
10
100
1k
10 k
100 k
1 M
10 M
40 M
3
GHz
-39.0 -78.7 -99.5 -124.4 -135.3 -137.5 -153.5 -164.5 -167.6
10
GHz
-34.6 -71.8 -91.1 -116.1 -124.1 -128.1 -147.1 -156.6 -160.0
18
GHz
-31.6 -66.5 -85.3 -110.3 -121.2 -125.3 -141.3 -150.7 -154.2
26.5
GHz
-29.0 -63.2 -81.8 -106.8 -117.8 -121.8 -137.8 -147.3 -150.8
E5052A with E5053A phase noise performance (standard,
correlation = 1, measuring ultra-low noise N5507A uW LO)
Block diagram of SSA millimeter wave
Figure 2-2. System phase noise sensitivity with the E5053A downconverter
(standard, 0 dBm input, correlation = 1, start offset frequency = 1 Hz, measurement time = 17.7 sec)