Atec Agilent-8564E User Manual

Page 11

Advertising
background image

11

Calibration interval
8560E, 8561E, 8562E, 8563E

2 years

8564E, 8565E

1 year

Warm-up time

5-minutes in ambient
conditions

Temperature

–10 °C to +55 °C (oper-
ating); –51 °C to +71 °C
(not operating)

Humidity

95% @ 40 °C for 5 days

Rain resistance

Drip-proof at 16 liters/
hour/sq. ft.

Altitude

15,000 ft. (operating),
50,000 ft. (non-
operating)

Pulse shock (half sine)

30 g for 11 ms duration

Transit drop

8-inch drop on six
faces and eight
corners

Electromagnetic compatibility

Conducted and radiated
interference in
compliance with
CISPR Pub. 11 (1990).
Meets Mil-STD-461C,
part 2, with certain
exceptions.

Power requirements

115 VAC operation:
90 to 140 V rms,
3.2 A rms max,
47 to 440 Hz
230 VAC operation:
180 to 250 V rms,
1.8 A rms max,
47 to 66 Hz

Maximum power dissipation
8560E, 8561E, 8562E, 8563E

180 W

8564E, 8565E

260 W

Audible noise (nominal)

<5.0 Bels power at
room temp (ISO DP7779)

Dimensions (w/o handle, cover)

337 mm W x 187 mm H

x 461 mm D

Weight (nominal)
8560E, 8561E, 8562E, 8563E

20 kg (44 lbs)

8564E, 8565E

21 kg (46 lbs)

Environmental specifications, Agilent 8560 E-series

Per MIL-T-28800, type III, class 3

3

, style C

Option 002 built-in tracking generator
(8560E only), continued

Spurious output (at +1 dBm output power)
Harmonic spurious

–25 dBc

Non-harmonic spurious

300 kHz to 2.0 GHz

–27 dBc

2.0 GHz to 2.9 GHz

–23 dBc

LO feedthrough

–16 dBm
(3.9 GHz to 6.8 GHz)

Residuals (RF-Power-Off)

–78 dBm
(300 kHz to 2.9 GHz

Dynamic range
TG feedthrough

1

300 kHz to 1 MHz

–95 dBm

1 MHz to 2.7 GHz

–115 dBm

2.7 GHz to 2.9 GHz

–110 dBm

Dynamic range

2

300 kHz to 1 MHz

96 dB

1 MHz to 2.7 GHz

116 dB

2.7 to 2.9 GHz

111 dB

Power sweep

10 dB range,
0.1 dB resolution

Inputs/outputs

RF output (front panel)

Type-N female, 50

Ω

(nominal)

Maximum safe reverse level

+30 dBm, ±30 Vdc

External ALC input

BNC female, use with

(rear panel)

negative detector

1. Leakage measured with maximum power into 50

Ω

and with 50

Ω

on RF input

2. Difference between maximum power output and tracking generator feedthrough
3. 8564E, 8565E: Class 5

Advertising