Atec Agilent-8568b-8566b User Manual
Page 5

5
CISPR EMI Amplitude
Parameter
Specification
Measurement range
+137 dBµV to noise floor (refer to amplitude sensitivity table)
Amplitude accuracy8
(+10° to +30°C, 0 – 20 dB RFP atten.)
±2.0 dB for 10 dB SA atten.
± 2.5 dB for 0 dB SA atten.
Calibration10
Automatic calibration for absolute amplitude with comb
generator (RFP)
Linearity check
3 dB input attenuator test for system overload and distortion
(RFP)
Impulsive signal tolerance
Input #1 withstands LISN transients (RFP)
Input #2 withstands CISPR calib. pulses (RFP)
Detection9
Peak and average (SA), quasi-peak (QPA)
Audio output
Speaker with volume control (QPA)
CISPR EMI Amplitude sensitivity
EMI receiver
Specification
Frequency band
Resolution bandwidth (6 dB)
Average noise6
(RFP/SA atten. = 0 dB)
Quasi-peak noise
(RFP/SA atten. = 0 dB)
10 – 150 kHz (band A)
200 Hz
-12 dBµV4
-17 dBµV typical
-15 dBµV (nominal)
150 kHz – 1 MHz (band B)
9 kHz
+4 dBµV
-1 dBµV typical
0 dBµV (nominal)
1 – 30 MHz (band B)
9 kHz
-13 dBµV
-16 dBµV typical
-11 dBµV (nominal)
30 – 1000 MHz (band C, D)
120 kHz
-1 dBµV
-6 dBµV typical
0 dBµV (nominal)