Frequency specifications – Atec Agilent-E4408B User Manual
Page 5

5
Span = 0 Hz
4 ms - 4000 s
Span ≥ 100 Hz
4 ms - 4000 s
Span = 0 Hz
401
Span ≥ 100 Hz
401
Delayed trigger range
1 µs to 400 s
Sweep (trace) points
Range
Accuracy (Span = 0 Hz)
±1%
Trigger type
Free run, single, line, video, offset, delayed, external
Basic analyzer
Sweep time and trigger
Range
(–3 dB)
1 kHz - 5 MHz
a
(–6 dB EMI)
9 kHz, 120 kHz
With 1DR
b
(–3 dB)
Add 100 Hz, 300 Hz
(–6 dB EMI)
Add 200 Hz
1 to 300 Hz
1 kHz to 3 MHz
5 MHz
100 to 300 Hz
1 kHz to 5 MHz
Range
with 1DR
< 15:1 synchronously tuned four poles, approximately Gaussian
Video bandwidths (1-3-10 sequence)
30 Hz to 3 MHz
Adds 1, 3, 10 Hz for RBWs less than 1 kHz
±15%
±30%
Selectivity (60 dB/3 dB bandwidth ratio)
< 5:1 digital, approximately Gaussian
Range
Accuracy
±10%
Basic analyzer
Resolution bandwidths (1-3-10 sequence)
a.
For resolution bandwidths < 1 kHz or > 3 MHz, not compatible with the rms detector.
b.
Only available for spans < 5 MHz.
Frequency Specifications
(continued)