Specifi cations – Atec Agilent-N9344C User Manual

Page 3

Advertising
background image

3

Specifi cations

Specification

Supplemental information

Frequency

Frequency range

1 MHz to 20 GHz (tunable to 9 kHz)

AC coupled

Internal 10 MHz frequency reference accuracy

Aging rate

± 1 ppm/year

Temperature stability

± 1 ppm

Referenced to frequency reading at 25 °C.
Temperature varied at max. of 2 °C per
minute. Control voltage held at voltage
control range midpoint

Frequency readout accuracy with marker (start, stop, center, marker)

Marker resolution

(frequency span)/(sweep points - 1)

Uncertainty

± (frequency indication × frequency
reference uncertainty + 1% × span + 20%
× resolution bandwidth + marker resolution
+ 1 Hz)

Frequency reference uncertainty = (aging
rate × period of time since adjustment +
temperature stability)

Marker frequency counter

Resolution

1 Hz

Accuracy

± (marker frequency × frequency reference
uncertainty + counter resolution)

RBW/span ≥ 0.02; marker level to displayed
noise level > 25 dB; frequency offset 0 Hz

Frequency span

Range

0 Hz (zero span), 100 Hz to 20 GHz

Resolution

1 Hz

Accuracy

± (0.22% × span + span/(sweep
points - 1))

Nominal

SSB phase noise

Carrier offset

30 kHz

< -86 dBc/Hz, typical -89 dBc/Hz

20 to 30 °C

100 kHz

< -97 dBc/Hz, typical -99 dBc/Hz

Center frequency 500 MHz

1 MHz

< -115 dBc/Hz, typical -119 dBc/Hz

Resolution bandwidth (RBW)

-3 dB bandwidth

10 Hz to 3 MHz

1-3-10 sequence

Accuracy

± 5%, RBW = 10 Hz to 1 MHz

Nominal

± 10%, RBW = 3 MHz

Resolution filter shape
factor

< 5:1

Nominal; 60 dB/3 dB bandwidth ratio;
digital, Gaussian-like

Video bandwidth (VBW)

-3 dB bandwidth

1 Hz to 3 MHz

1-3-10 sequence

Accuracy

± 10%, VBW = 1 Hz to 1 MHz

Nominal

Advertising