Speed characteristics, Meas) – Atec Agilent-53200A Series User Manual
Page 11

11
53210A
53220A
53230A
Measurement/IO timeout
(nom)
no timeout or 10 ms to 2000 s, in 1 ms steps
Auto-level speed
Slow mode (50 Hz): 350 ms (typ)
Fast mode (10 kHz): 10 ms (typ)
Configure-change speed
Frequency, Period, Range, Level: 50 ms (typ)
Single measurement throughput
5
: readings/s
(time to take single measurement and transfer from volatile reading memory over I/O bus)
Typical (Avg. using READ?):
LAN (VXI-11)
110
120
LAN (sockets)
200
200
USB
200
200
GPIB
210
220
Optimized (Avg. using *TRG;DATA:REM? 1, WAIT):
LAN (VXI-11)
160
180
LAN (sockets)
330
350
USB
320
350
GPIB
360
420
Block reading throughput
5
: readings/s (Example uses: 50,000 readings)
(time to take blocks of measurements and transfer from volatile reading memory over I/O bus)
Typical (Avg. using READ?):
LAN (VXI-11)
300
990
8700
LAN (sockets)
300
990
9700
USB
300
990
9800
GPIB
300
990
4600
Optimized (Avg. using *TRG;DATA:REM? 1, WAIT):
LAN (VXI-11)
300
990
34700
LAN (sockets)
300
990
55800
USB
300
990
56500
GPIB
300
990
16300
Speed Characteristics
4
(meas)