Atec Rohde-Schwarz-ESU-40 User Manual
Page 9

Version 02.00, October 2007
R&S
®
ESU EMI Test Receiver
9
R&S
®
ESU26/40
20 MHz ≤ f < 30 MHz, BW = 9 kHz
<–16 dBµV
30 MHz ≤ f < 500 MHz, BW = 120 kHz
<–5 dBµV
500 MHz ≤ f < 1 GHz, BW = 120 kHz
<–3 dBµV
1 GHz ≤ f < 2 GHz, BW = 1 MHz
<6 dBµV
2 GHz ≤ f < 3 GHz, BW = 1 MHz
<7 dBµV
3 GHz ≤ f ≤ 3.6 GHz, BW = 1 MHz
<11 dBµV
max peak
typ. +11 dB
RMS
typ. +1 dB
quasi peak
band A
typ. +3 dB
band B
typ. +4 dB
Increase of DANL relative to AV display
bands C and D
typ. +6 dB
Immunity to interference
f ≤ 3.6 GHz
>90 dB suppression, typ. >110 dB
f > 3.6 GHz
>70 dB suppression, typ. >100 dB
Image frequency
f = receive frequency
f ≤ 3.6 GHz
>90 dB suppression, typ. >110 dB
3.6 GHz < f ≤ 4.2 GHz
typ. 70 dB suppression
f > 4.2 GHz
>70 dB suppression, typ. >90 dB
Intermediate frequency
f = receive frequency
Spurious response
f > 1 MHz, without input signal,
RF attenuation = 0 dB, termination = 50 Ω
<–103 dBm
∆f > 100 kHz
mixer level <–10 dBm, f
in
≤ 2.3 GHz
<–80 dBc
mixer level <–35 dBm, 2.3 GHz <f
in
<4 GHz <–70 dBc
mixer level <–10 dBm
4 GHz ≤ f < 8 GHz
<–70 dBc
8 GHz ≤ f < 16 GHz
<–64 dBc
16 GHz ≤ f < 26 GHz
<–58 dBc
26.5 GHz ≤ f < 40 GHz
<–52 dBc
Other interfering signals
f = receive frequency
Level display (analyzer mode)
Screen
625 × 500 pixels (one diagram), max.
2 diagrams with independent settings
Logarithmic level axis
1 dB to 200 dB, in steps of 1/2/5
Linear level axis
10 % of reference level per level division,
10 divisions or logarithmic scaling
1 measurement diagram
3
Number of traces
2 measurement diagrams
6
Trace detector
max peak, min peak, auto peak (normal),
sample, RMS, average, quasi peak,
CISPR-AV, CISPR-RMS
default value
625
Number of measurement points
range
155 to 30001 in steps of about a factor of 2
Trace functions
clear/write, max hold, min hold, average
local measurement, display update rate,
625 points, zero span
80 per second
remote measurement, display off:
zero span/sweep time 1 ms
70 per second
Trace update rate
span = 10 MHz, sweep time 2.5 ms
50 per second
Setting range of reference level
logarithmic level display
–130 dBm to (+5 dBm + RF attenuation),
max. 30 dBm, in steps of 0.1 dB
linear level display
7.0 nV to 7.07 V in steps of 1 %
logarithmic level display
dBm, dBµV, dBmV, dBµA, dBpW
Units of level axis
linear level display
µV, mV, µA, mA, pW, nW