Portable, rugged, and ergonomic, Ef h, Physical features – Atec Panametrics-Olympus-Epoch-600 User Manual

Page 7: Standard package, Physical features – rear

Advertising
background image

7

E

F

H

Physical Features

The EPOCH 600 is a lightweight, portable flaw detector built to
be rugged and flexible for nearly any inspection. Some key physi-
cal features include:

•

Large, full VGA resolution transflective display for vivid clarity

in any indoor lighting and direct sunlight conditions

•

Rubber overmolded bumpers on all four corners for shock

absorption and anti-marring considerations

•

Four-point harness connection for chest straps

•

“No tools” access to battery compartment and side I/O door

•

Continuous position stand with right angle crook for improved

stability from 0 to 180 degrees

•

Gasketed side door for USB OTG connection and removable

memory

•

Standard internal, rechargeable lithium-ion battery

•

Standard AA battery tray for extended portable use

•

Lightweight, ergonomic design for increased portability and

ease of use

Portable, Rugged, and Ergonomic

Standard Package

•

EPOCH 600 Digital ultrasonic flaw detector, AC or battery

operation

•

Charger/AC adapter (100 VAC, 115 VAC, 230 VAC ,

50 Hz or 60 Hz)

•

Rechargeable lithium-ion battery

•

AA battery tray

•

Transport case

•

Basic operation user’s manual

•

Quick reference card

•

Comprehensive operation manual (CD)

Physical Features – Rear

A – USB On-The-Go port
B – MicroSD card
C – DC power connector
D – VGA port
E – RS-232/Alarm port
F – Transducer connectors (2)
G – Battery door
H – Pipe stand

The EPOCH Flaw Detector Family: The EPOCH flaw detector line
features vertical and horizontal layout instruments that span the
range of inspection capabilities from basic to advanced.
The EPOCH 600 and EPOCH 1000 Series flaw detectors feature
horizontal layout instruments for enlarged A-scan and image
representation in a portable format, and provide an exceptionally
high quality of digital flaw detection technology.

Instrument Inputs/Outputs

USB ports

USB On-The-Go (OTG)

RS-232 port

Yes

Video output

VGA output standard

Analog output

1 analog output (optional), Selectable 1 V/10 V Full

Scale, 4 mA max

Digital output

3 alarm outputs, 5 V TTL, 10 mA

Environmental Ratings

IP rating

Designed to meet the standards of the Ingress Protec-

tion (IP) rating number IP67 (navigation pad version)

or IP66 (knob version) per IEC 60529-2004 (Degrees

of Protection provided by enclosures – IP Code).

The product design was confirmed to meet the IP

rating by means of Olympus NDT’s internal design

verification test process that occurs prior to the release

of the product to production.

Explosive atmosphere

Safe operation as defined by Class I, Division 2,

Group D, as defined in the National Fire Protection

Association Code (NFPA 70), Article 500, and tested

using MIL-STD-810F, Method 511.4, Procedure I.

Shock tested

MIL-STD-810F, Method 516.5 Procedure I,

6 cycles each axis, 15g, 11 ms half sine.

Vibration tested

MIL-STD-810F, Method 514.5, Procedure I, Annex C,

Figure 6, general exposure: 1 hour each axis

Operating temperature

–10 °C to 50 °C (14 °F to 122 °F)

Battery storage temperature 0 °C to 50 °C (32 °F to 122 °F)

Advertising