Measurement speed – Atec Rohde-Schwarz-ZNB Series User Manual
Page 6

Version 07.00, August 2013
6
Rohde & Schwarz R&S
®
ZNB Vector Network Analyzer
Measurement speed
Measurement time
for 201 measurements points, with 200 MHz span, 500 kHz measurement bandwidth
with 900 MHz center frequency
< 5 ms
with 5.1 GHz center frequency
< 4 ms
Measurement time per point
500 kHz measurement bandwidth,
CW mode
2.5 µs
Sampling time per point
at 1 MHz measurement bandwidth
860 ns
at 10 MHz measurement bandwidth
312 ns
Data transfer time
for 201 measurements points
via IEC/IEEE bus
typ. 2.5 ms
via VX11 over 1 Gbit/s LAN
typ. 1.6 ms
via RSIB over 1 Gbit/s LAN
typ. 1.0 ms
Time for measurement and data transfer
for 201 measurements points, with
800 MHz start frequency, 1 GHz stop
frequency, 1 MHz measurement
bandwidth (no additional time for data
transfer is needed, as this occurs
simultaneously during the measurement)
typ. 3 ms
Switching time between channels
with no more than 2001 points
< 5 ms
Switching time between two preloaded
instrument settings
with no more than 2001 points
< 5 ms