General specifications, Safety specification, Battery information – Atec Fluke-80i-110S User Manual

Page 2: Instrument compatibility, Warranty

Advertising
background image

Fluke Corporation
PO Box 9090, Everett, WA USA 98206

Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands

For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com/

Fluke. Keeping your world

up and running.

Floating voltage (output
cable and connector to
ground):
600 V ac rms on Installation
Category II per IEC 1010-1
circuits.
300 V ac rms on Installation
Category III per IEC 1010-1
circuits.
Basic accuracy (dc to 1 kHz)

Extended accuracy:
For other frequencies, refer to
the appropriate input current
range and add the error listed
below to the “Basic Accuracy.”

Phase shift, dc to 65 Hz:
10 A range < 1.5 degree
100 A range < 1 degree
Input load impedance
(of host instrument):
> 1 MΩ in parallel with up to

100 pF
Useful bandwidth (-3 dB):
0 to 100 kHz
Rise or fall time:

< 4 msec

Output noise level:
10 mV/A typ. 480 mV pk-pk
100 mV/A typ. 3 mV pk-pk
Maximum non-destructive
current:
0 to 2 kHz 140 A peak
2 to 10 kHz 110 A peak
10 to 20 kHz 70 A peak
20 to 50 kHz 30 A peak
50 to 100 kHz 20 A peak
Temperature coefficient:
2000 ppm/°C max. for tempera-
ture from 0 °C to 50 °C (32 °F to
132 °F)

All Electrical Specifications are valid at a
temperature of 3 °C to 23 °C (5 °F to 73 °F)

General specifications

Dimensions:
67 x 231 x 36 mm
(2.6 x 9.1 x 1.4 inches)
Weight:

330 g (11.6 oz), battery

included
Output cable:
1.6 meters (63 inches)
Maximum conductor size:
11.8 mm (.46 inch)
Maximum jaw opening:
12.5 mm (.49 inch)
Temperature:
Operating: 0 °C to 50 °C
(32 °F to 122 °F)
Non-operating: -30 °C to 70 °C
(-22 °F to 158 °F)
Relative humidity (operating):
0 to 85 % (0 °C to 35 °C; 32 °F
to 95 °F)
0 to 45 % (35 °C to 50 °C;
95 °F to 122 °F)
Altitude:
Operating: 0 to 2,000 meters
(0 to 6,560 feet)
Non-operating: 0 to 12,000
meters (0 to 40,000 feet)

Safety specification

Designed to meet the require-
ments of IEC 1010 and CSA-
C22.2 No. 1010.1: Installation
Category II, Working Voltage
600V, Pollution Degree 2.
Installation Category III,
Working Voltage 300V,
Pollution Degree 2.

Installation (Overvoltage)

Category II refers to local
level, appliances, and portable
equipment.

Installation (Overvoltage)

Category III refers to distribu-
tion level and fixed installation
circuits inside a building
electrical service entrance.

Designed to meet the

requirements of UL1244,
Protection Class II Double
Insulation.

Battery information

Battery:

9 volt, IEC 6LR61

Consumption:

8.6 mA typical;

12 mA maximum
Service life:
55 hours typical, when Alkaline
IEC 6LR61 is used
40 hours minimum, when
Alkaline IEC 6LR61 is used

Battery indicator (ON):
Green LED dims when battery
voltage is below 6.5V
Overload indicator (OL):
Red LED indicates that wave-
form or impulse is out of range

Instrument compatibility

The 80i-110s is compatible
with any Fluke ScopeMeter Test
Tool, Power Harmonics
Analyzer, Oscilloscope,
Multimeter, or other voltage
measurement device that has
the following features:

BNC input connector. A
BNC-to-banana adapter
(order PM 9081/001 from
Fluke) can also be used with
standard inputs on a digital
multimeter (DMM).

Input accuracy of 2 % or
better to take full advantage
of the accuracy of the probe.

Input impedance of greater
than or equal to 1 MΩ in

parallel with a maximum of
100 pF.

A passband of more than
four times the frequency
of the waveform to be
measured.

Warranty

One-year warranty

Input current

Error (after zero check)

dc or ac peak

100 mV/A

10 mV/A

50 mA to 10 A

< 3 % of reading

+ 50 mA

50 mA to 40 A

< 4 % of reading
+ 50 mA

40 to 80 A

< 12 % of reading
+ 50 mA

80 to 100 A

< 15 % of reading

Additional error

Frequency

100 mV/A

10 mV/A

1 to 5 kHz

3 %

3 %

5 to 20 kHz

12 %

12 %

> 20 kHz

Not specified

Not specified

Ordering information

Model
80i-110s

AC/DC Current Probe

Page 2 of 2, Document #2228
80i-110s AC/DC Current Probe
©2002 Fluke Corporation
Rev. A-9/2002

Advertising