Specifications – Atec Fluke-Norma-4000-5000 Series User Manual

Page 5

Advertising
background image

ENEN

Ambient conditions
Working Temperature Range

5 °C to 35 °C (41 °F to 95 °F)

Storage Temperature Range

-20 °C to 50 °C (-4 °F to 122 °F)

Housing

Equipped with a solid metal case to meet stringent EMC requirements.

Weight

Fluke Norma 4000: Base unit approx. 5 kg (11 lb.)

Fluke Norma 5000: Base unit approx. 7 kg (15 lb.)

Size (HxWxD)

Fluke Norma 4000:

15 cm x 23.7 cm x 31.5 cm (5.9 in. x 9.3 in. x 12.4 in.)

Fluke Norma 5000:

15 cm x 44.7 cm x 31.5 cm (5.9 in. x 17.6 in x 12.4 in.)

Display

5.7“ / 144mm - 320 x 240 pixel

User-selectable background lighting and contrast.

Climatic Class

KYG DIN 40040, max. 85 % relative humidity,

non-condensing.

Net Connection

85 to 264 V ac, 50 to 60 Hz, dc 100 to 260 V, ca. 40 VA European plug with switch.

Measuring Inputs

Safety sockets 4 mm, 2 for each input. External shunt connection over BNC socket.

Operation

Membrane keyboard with cursor – function keys and direct functions.

Connections

Rear panel of the 3-phase analyzer

Measured Values

Non-gapping calculation of averaged values for each phase. In three phase system additionally calculation of total

power and averaging of V and I of the three phases. The fundamental H01 will be calculated in synchronous mode

also for these values.
U

RMS

effective value, U

rm

rectified mean, U

m

mean value

U

p

-, U

p

+, U

pp

peak values

U

cf

crest factor U

cf

, U

ff

form factor

U

fc

fundamental content

U

thd

distortion factor DIN, IEC

I

RMS

effective value, I

rm

rectified mean, I

m

mean value

I

p-

, I

p+

, I

pp

peak values

I

cf

crest factor I

cf

, I

ff

form factor

I

fc

fundamental content

I

thd

distortion factor DIN, IEC

P active power [W]
Q reactive power [Var]
S apparent power [VA]
ë, cos. phase angular
Integral function for active power P, reactive power Q, apparent power S, voltage (U

m

) and current (I

m

),

Number of digits 4 or 5 dependent on measured value.

Frequency and Synchronization
Range

DC and 0.2 Hz to sampling rate

Accuracy

±0.01 % of measured value (reading)

•

Channels which can be selected: all U/I or external input.

•

One of three low pass filter with different frequencies can be switched into the signal.

•

The frequency is always visible on the top of the screen.

•

The BNC synchronization socket on backside of the instrument can be used either as input or output.

•

The input signals can be measured up to the sample rate of the power phase. The maximum level must not be higher then 50 V.

•

The output signal is a pulsed 5 Volts TTL signal (frequency depends on the measured synch frequency).

Configuration Memory
Up to 15 user configurations can be saved into a permanent memory and reloaded later on. Changes that were not saved are lost after switching off the instrument.

Interface
RS232/USB interface for upload of firmware and data exchange with the PC. A printer can be connected over an external converter.
Options

IEEE 488.2 / 1 MBit/s

Ethernet / 10 MBit/s or 100 Mbit/s

Standards and Safety
Electrical Safety

EN 61010-1 / 2

nd

Edition 1000 V CAT II (600 V CAT III)

Degree of pollution 2, safety Class I

EN 61558 for transformer

EN 61010-2-031/032 for accessories

Maximum Inputs

For voltage inputs Measurement range 1000 V

eff

, 2 kV

peak

For current inputs Measurement range 10 A

eff

, 20 A

peak

Test Voltages

Net input - case (protective conductor): 1.5 KV ac

Net connection – Measurement input: 5.4 kV ac

Measurement inputs – Case: 3.3 kV ac

Measurement input – Measurement input: 5.4 kV

Electromagnetic Susceptibility

Emission: IEC 61326-1, EN 50081-1, EN 55011 Class B

Immunity: IEC 61326-1 / Annex A (industrial sector), EN 50082-1

5

Specifications

Fluke Corporation Fluke Norma 4000/5000 power analyzers