Transmit modulation – Atec Rohde-Schwarz_FSL Series User Manual
Page 22

R&S
®
FSL Spectrum Analyzer
Version 09.00, December 2008
22
Standard deviation
σ of spectrum emission mask measurement as a function of sweep time (T
sweep
)
Transmit modulation
Composite EVM
test case 6.7.1
Measurement range
1.0 % to 25 %
Inherent EVM
<1.9 %
Measurement uncertainty
test models 1 to 5
P > –40 dBm
<1 % (
σ = 0.3 %)
PMU: <2.5 %
Peak code domain error power
(PCDEP)
test case 6.7.2
Measurement range
–50 dB to 0 dB
0 dB to –50 dB
Inherent PCDEP
<
−50 dB (σ = 0.95 dB)
–30 dB < PCDEP
<0.15 dB (
σ = 0.05 dB)
PMU: <1.0 dB
–40 dB < PCDEP < –30 dB
<0.15 dB (
σ = 0.05 dB)
PMU: <1.0 dB
–50 dB < PCDEP < –40 dB
<0.15 dB (
σ = 0.05 dB)
PMU: <1.0 dB
Measurement uncertainty
–60 dB < PCDEP < –50 dB
<0.15 dB (
σ = 0.05 dB)
PMU: <1.0 dB
10
2
10
3
0
0.5
1
1.5
2
2.5
T
sw eep
in ms
S
tan
da
rd
de
v
iat
ion
in
d
B
relative (RBW : 30kHz)
absolute (RBW : 30kHz)