Main Output

The main output on the unit has four taps, allowing the selection

of output voltages up to 240V and output currents up to 200A.

Range Continuous 5 minutes 1 minute 6 sec.

10V 25A 50A 100A 200A*
35V 10A 20A 30A
100V 3A 5A 10A
240V 1A 2A 3A
240V DC 1A 2A 3A

The above intermittent on times must be followed by an off time

of 15 minutes, and are based on an ambient temperature of

25°C.

*6 second intermittent ratings available with 230V supply.

Auxiliary AC and DC Outputs

A switched, isolated auxiliary DC supply is available to power the

relay under test. The output is protected by a fuse and

electronic trip circuit.

DC stabilised output 24V 1.04A
48V 0.52A
60V 0.42A
110V 0.23A
220V 0.11A

Variable AC output 0-130Vac 300mA
The variable AC output is isolated, switched, and fuse protected.

Metering

The output is metered by a digital true RMS system with a

memory ammeter - whenever the timer stops and the output is

switched off, the current reading is held on the display.

Range Resolution Trip current Accuracy

5.000A 0.001A 5.5A ±0.5%rdg+5d
20.00A 0.01A 22A ±0.5%rdg+5d
50.00A 0.01A 55A ±0.5%rdg+5d
200.0A 0.1A 220A ±0.5%rdg+5d

A current trip is automatically set to 110% of full scale of the

selected metering range to protect the device under test.

Auxiliary Metering Inputs

An auxiliary metering input is provided which is able to measure

RMS and DC voltage or current. In addition the frequency of the

external input may be measured, and the phase measured

between the main output current and the auxiliary metering input.

Setting Range Resolution Accuracy

Volts (DC) 270.0V 0.1V ±0.7%rdg+5d
Volts (AC) 270.0V 0.1V ±0.7%rdg+5d
Amps (DC) 5.000A 1mA ±0.7%rdg+5d
Amps (AC) 5.000A 1mA ±0.7%rdg+5d
Phase 0.1° ±3°
Frequency 20-100Hz 0.1Hz ±0.2%rdg+1d
The current input is protected by a F6.3A fuse.

Supply Requirements

115V±10% 50/60Hz 1ph 2300VA max

230V±10% 50/60Hz 1ph 2300VA max

The correct range is automatically selected by the unit.

Timing System

Range 0-999.999s
Resolution 1ms
Accuracy 0.1%rdg+2d (+4d current operated mode)

The contact circuit has an open circuit voltage of 24Vdc and a

short circuit current of 10mA. Each contact circuit will auto -

select for normally open or normally closed contacts. A DC

voltage of 24-240Vdc may also be used to trigger either timer

channel. The output is automatically switched off at the end of

the test to safeguard the relay under test.

The following functions are provided:

Mode Timer Start Timer Stop

Internal start Press ‘ON’ Contact 1 or 2
Single contact Contact 1 Contact 1
Dual contact Contact 1 Contact 2
Current operated Current > 10% Current < 10%
of range of range

Pulse mode is used for setting the current level in devices

sensitive to heating, and allows current to be injected for 400ms

and the current recorded.

RS232

An RS232 port is provided to allow connection of a printer.

Protection and Safety

The unit is protected by electronic over current, duty cycle trips

and thermal monitoring on the outputs and fuses on the input

and regulator. The auxiliary AC output is fuse protected, and the

stabilised DC output is protected by an electronic trip and fuse.

An earth terminal is provided for connection to a local earth. The

unit is designed to comply with BSEN61010, and is CE marked.

Temperature Range

Storage -20°C to 60°C Operating 0°C to 45°C

Lead Set specifications

The 200ADM is supplied with a lead set including:

2 x 5m 25mm

2

200A leads terminated in M10 ring crimps

2 x 5m 2.5mm

2

25A leads terminated in 4mm plugs

1 x 5m 2 core auxiliary leads terminated in 4mm plugs
The lead set is supplied in a plastic carrying case.

Dimensions Weight

380mm x 314mm x 221mm 20.7kg

Including handle and corner protectors

Accessories

Operating manual, o/p lead set, mains lead, spare fuses

Optional Accessories

Filter unit, Printer, 200ADM-DRE (see below), pushbutton lead

for runback timing on disc induction relays.

Optional 200ADM-DRE extension unit

The 200ADM-DRE converts the auxiliary

ac output into a 0-5A 6.8V or 0-10A 3.4V

current output, allowing bias differential

relays to be tested. The unit may also be

used in other situations where a second

current output is required from the 200ADM.

200ADM Specification