Characteristics, Data sheet – Atec Tektronix-TCPA Series User Manual

Page 2

Advertising
background image

Data Sheet

Meets Today's AC/DC Current Measurement
Applications

The TCPA300 amplifier, when used with TCP312, TCP305, or TCP303
probes, provides a wide range of current measurement capability and spans
the gap between low-level milliamp measurements to very high current
levels. These three probes provide current measurement capabilities of
30 A, 50 A, and 150 A DC continuous. For even higher current levels, the
TCPA400 amplifier with the TCP404XL current probe measures 500 A DC
continuous and 750 A DC continuous, derated with duty cycle.
Higher frequency performance is available with the TCP312 w/TCPA300
providing ≥100 MHz bandwidth and a maximum current of 30 A DC.

Measurement Errors and Manual Calculations are Now
a Thing of the Past

With this new series of current measurement tools, automatic control and
on-screen scaling and units is provided for users of Tektronix TDS3000,
TDS500, TDS600, TDS700, TDS5000, TDS6000, and TDS7000B Series
oscilloscope systems (the DPO3000, DPO4000, and DPO7000 Series
oscilloscopes, the TPA-BNC adapter is required).
The TCP300/TCP400 current measurement systems seamlessly integrate
with your TDS Series oscilloscope.
Even non-TEKPROBE systems can use the TCPA300/400 Series to make
proper current measurements by simply multiplying the measured output
voltage on the oscilloscope by the TCPA300/400 Series range setting.

Characteristics

Characteristic

TCP312 w/ TCPA300

TCP305 w/ TCPA300

TCP303 w/ TCPA300

TCP404XL w/ TCPA400

Bandwidth (–3 dB)

DC – 100 MHz

DC – 50 MHz

DC – 15 MHz

DC – 2 MHz*

3

Rise Time

≤3.5 ns

≤7 ns

≤23 ns

≤175 ns

Maximum Current Ratings:
High-current Sensitivity Range

10 A/V Range

10 A/V Range

50 A/V Range

1 A/mV Range

DC (Continuous)

30 A

50 A

150 A

500 A (750 A*

4

)

RMS (Sinusoidal)

21.2 A

35.4 A

150 A

500 A

Peak Pulse

50 A

50 A

500 A

750 A

Low-current Sensitivity Range

1 A/V Range

5 A/V Range

5 A/V Range

N/A

DC (Continuous)

5 A

25 A

25 A

N/A

RMS (Sinusoidal)

3.5 A

17.7 A

17.7 A

N/A

Peak Pulse

50 A

50 A

500 A

N/A

DC Accuracy

(Operating temp 0 °C to 50 °C)

±3% of reading

±3% of reading

±3% of reading (10 °C to 50 °C)

+3%/–6% of reading (0 °C

to <10 °C)

±3% of reading

DC Accuracy, Typical

(Operating temp 23 °C ±5 °C)

±1% of reading

±1% of reading

±1% of reading

±1% of reading

Nominal

Maximum Bare Wire Voltage

For use with insulated wires only

600 V

RMS

CAT I & II

300 V

RMS

CAT III

1 mA

5 mA

5 mA

1 A

Lowest Measurable Current

(at ±3% accuracy at DC)

Scope set to 1 mV/div and 20 MHz BW Limited

Scope set to 1 mV/div and 20 MHz BW Limited

Insertion Impedance

(See curves below)

0.08 Ω at 1 MHz

0.15 Ω at 10 MHz

0.27 Ω at 50 MHz

0.7 Ω at 100 MHz

0.035 Ω at 1 MHz

0.12 Ω at 10 MHz

0.4 Ω at 50 MHz

0.01 Ω at 1 MHz

0.025 Ω at 5 MHz

0.1 Ω at 15 MHz

0.1 mΩ at 10 kHz

0.6 mΩ at 100 kHz

8 mΩ at 1 MHz

16 mΩ at 2 MHz

2

www.tektronix.com