Datasheet

Characteristics

Warranted

Characteristic

Description

Bandwidth

DC to ≥20 MHz (-3 dB). Derate with Frequency. See

Figure 1. (Calculated from Rise Time BW = .35/Rise

Time)

Rise Time

≤17.5 ns

Max RMS Current

150 A

Max Peak Pulse Current

500 A

Minimum Sensitivity

5 mA (on oscilloscope’s 1 mV/div setting)

AC Coupling (on oscilloscopes that support AC/DC

coupling)

Current Ranges

25 A and 150 A

Max Bare-wire Voltage

600 V

RMS

CAT I and II

300 V

RMS

CAT III

Typical

Characteristic

Description

DC Accuracy

±1% Typical

3% Warranted

Max Amp-second Product

25 A Range

3,000 Aµs

150 A Range

15,000 Aµs
0.03 Ω at 1 MHz
0.075 Ω at 5 MHz
.125 at 10 MHz

Insertion Impedance

.3 at 20 MHz

Signal Delay

≈ 21 ns

Displayed RMS Noise

≤500 μA

RMS

(probe only)

Environmental

Characteristic

Description

Temperature

Operating

0 ºC to +50 ºC

Nonoperating

-40 ºC to +75 ºC

Humidity

Operating

5% to 95% RH (Relative Humidity) at up to +30 ºC

5% to 85% RH above 30 ºC up to +50 ºC,

noncondensing

Nonoperating

5% to 95% RH (Relative Humidity) at up to +30 ºC

5% to 85% RH above 30 ºC up to +75 ºC,

noncondensing

Altitude

Operating

Up to 2,000 meters (6,560 feet)

Nonoperating

Up to 12,000 meters (40,000 feet)

Physical Characteristics

Dimension

Meters

Inches

Cable Length

2 meters

79 in.

Probe head size

Length

26.8 cm

10.55 in.

Width

4.1 cm

1.6 in.

Height

15.6 cm

6.13 in.

Max Conductor Diameter

21 mm x 25 mm

0.83 in. x 1.0 in.

Weight

kg

lb.

Shipping

.7865

1.734

Power Requirements

TCP0150 is powered directly by oscilloscopes

with the TekVPI interface

Recommended Oscilloscopes
Oscilloscopes with the TekVPI probe interface.
Note: For best probe support, download and install the latest version of the
oscilloscope software from www.tektronix.com