Atec Tektronix-TDS5000B Series User Manual
Page 7

Digital Phosphor Oscilloscopes
•
www.tektronix.com/tds5000b
7
Digital Phosphor Oscilloscopes
TDS5000B Series
A c q u i s i t i o n S y s t e m
TDS5032B/5052B
TDS5034B/5054B/5104B
TDS5054BE
Real-time Sample Rates
1 Channel (max)
5 GS/s
5 GS/s
1 GS/s
2 Channels (max)
2.5 GS/s
2.5 GS/s
1 GS/s
3-4 Channels (max)
—
1.25 GS/s
1 GS/s
Equivalent Time Sample Rate (max)
250 GS/s
250 GS/s
200 GS/s
Maximum Record Length per Channel
8M/4M/—
8M/4M/2M
8M/4M/2M
with Standard Memory
With Opt. 3M
16M/8M/—
16M/8M/4M
16M/8M/4M
M a x i m u m D u r a t i o n a t H i g h e s t R e a l - t i m e R e s o l u t i o n ( 1 c h )
TDS5054BE
All Other Models
Time Resolution (single shot)
1 ns (1 GS/s)
200 ps (5 GS/s)
Max Duration with Standard Memory
8 ms
1.6 ms
Max Duration with Opt. 3M
16 ms
3.2 ms
A c q u i s i t i o n M o d e s
All Models
FastAcq Acquisition
FastAcq optimizes the instrument for analysis of dynamic signals and capture
of infrequent events. Maximum FastAcq waveform capture rate
is 100,000 wfms/s
Sample
Acquire sampled values
Peak Detect
Captures narrow glitches (<1 ns) at all real-time sampling rates
Averaging
From 2 to 10,000 waveforms included in average
Envelope
From 2 to 2 x 109 waveforms included in min-max envelope
Hi-Res
Real-time boxcar averaging reduces random noise and increases resolution
Waveform Database
Accumulates a waveform database that provides a three dimensional array
of amplitude, time, and counts
FastFrame
™
Acquisition
Acquisition memory divided into segments; maximum trigger
rate >225,000 waveforms per second