Atec Tektronix-VM100-VM101 User Manual
Page 3

Appendix A: Specifications
VM 100 & VM101 User Manual
A–3
Table A–3: Measurement accuracy
Category
1
Description
Differential Phase
REQ:
<
1.0
°
RI: Typically
<
0.5
°
from 20
°
C to 30
°
C with test signal generator source
Differential Gain
REQ:
<
2.0%
RI: Typically
<
1% from 20
°
C to 30
°
C
for nominal 300 mV chrominance
amplitude
Chrominance-to-Luminance Gain
REQ:
<
2%
RI: Typically
<
1% from 20
°
C to 30
°
C
with test signal generator source
Chrominance-to-Luminance Delay
REQ:
<
20 ns
REQ:
<
10 ns from 20
°
C to 30
°
C with test
signal generator source.
RI: Typically
<
5 ns from 20
°
C to 30
°
C with test signal generator source
Frequency Response
REQ:
<
0.2 dB
RI: Typically
<
0.1 dB (1 %) from 20
°
C
to 30
°
C with test signal generator
source
RI: VM100: 50 kHz to 4.2 MHz
VM101: 50 kHz to 5.8 MHz
2T K-Factor
REQ:
2 %
RI:
Typically
<
0.5 % from 20
°
C to 30
°
C with test signal generator source
1
All tests are performed with AVG enabled. Processing time for each
measurement is a function of the signal-to-noise ratio.